Difference between revisions of "Lab Manual: Optical Microscopy"

From Course Wiki
Jump to: navigation, search
(Estimating the diffusion coefficient by tracking suspended microspheres)
m (Introduction)
 
(286 intermediate revisions by 8 users not shown)
Line 1: Line 1:
 
[[Category:Lab Manuals]]
 
[[Category:Lab Manuals]]
 
[[Category:20.309]]
 
[[Category:20.309]]
<div style="padding: 10px; width: 7.5in; border: 3px solid #000000;">
+
[[Category:Optical Microscopy Lab]]
 +
{{Template:20.309}}
 +
[[Image:Hooke-CorkMicrograph.png|center|300px|Hooke micrograph of cork cells]]
  
==Introduction==
+
<blockquote>
In this lab, you will design and build an optical microscope from optical components.  
+
<div>
 +
''I took a good clear piece of Cork, and with a Pen-knife sharpen'd as keen as a Razor, I cut a piece of it off, and thereby left the surface of it exceeding smooth, then examining it very diligently with a Microscope, me thought I could perceive it to appear a little porous; but I could not so plainly distinguish them, as to be sure that they were pores, much less what Figure they were of: But judging from the lightness and yielding quality of the Cork, that certainly the texture could not be so curious, but that possibly, if I could use some further diligence, I might find it to be discernable with a Microscope, I with the same sharp Penknife, cut off from the former smooth surface an exceeding thin piece of it, and placing it on a black object Plate, because it was it self a white body, and casting the light on it with a deep plano-convex Glass, I could exceeding plainly perceive it to be all perforated and porous, much like a Honey-comb, but that the pores of it were not regular; yet it was not unlike a Honey-comb in these particulars.''
  
By adding components, you can add additional capabilities such as dark field, confocal, or super-resolution microscopy.
+
''I told several lines of these pores, and found that there were usually about threescore of these small Cells placed end-ways in the eighteenth part of an Inch in length, whence I concluded there must be neer eleven hundred of them, or somewhat more then a thousand in the length of an Inch, and therefore in a square Inch above a Million, or 1166400. and in a Cubick Inch, above twelve hundred Millions, or 1259712000. a thing almost incredible, did not our Microscope assure us of it by ocular demonstration.''
  
===Objectives===
+
<blockquote>
* Learn about the theory and practice of light microscopy
+
''&mdash; [http://en.wikipedia.org/wiki/Robert_Hooke Robert Hooke] from Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon (1665)<ref name="Micrographia">Hooke, R.  [http://www.gutenberg.org/files/15491/15491-h/15491-h.htm Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon] London:Jo. Martyn, and Ja. Allestry, Printers to the Royal Society; 1665</ref> ''
* Use ray tracing rules to design a transmitted bright field and fluorescent light microscope
+
</blockquote>
* Construct the microscope from optical components
+
</div>
* Characterize the microscope's performance
+
</blockquote>
* Record, enhance, and analyze microscope images
+
<br/>
* Track moving particles
+
* Make quantitative measurements of biological systems
+
  
===How to do this lab===
+
==Introduction==
====Week 1====
+
In this lab, you will build an optical microscope using lenses, mirrors, filters, optical mounts, CCD cameras, lasers, and other components in the lab. The work is divided into 3 parts. Each part requires some lab work, some analysis, lots of clear thinking, and a written report. You will submit a short, group report after parts 1 and 2. The final report should include results from all 3 parts of the lab. You may revise and improve your part 1-2 reports before the final submission.
* Read the references; understand lenses, ray tracing, and magnification
+
* Design and build microscope
+
* Characterize the transmitted bright field performance of the microscope
+
* Add a laser illumination beam path
+
  
====Week 2====
+
===Part 1===
* Characterize the fluorescent imaging performance of the microscope
+
[[Image:Hooke-Microscope.png|thumb|Robert Hooke's microscope]]
* Image fluorescent samples
+
In part 1 of the lab, you will build a compound microscope, determine its magnification, and attempt to measure the size of microscopic objects. The instrument you create will have a great deal in common with the microscope Robert Hooke built in the mid-1660s. Hooke meticulously documented his microscopic observations and published them in a popular volume called ''Micrographia'' in 1665. The measurements you make in part 1 will call to mind Hooke's early quantification of the size of plant cells (see quote at top of page). You will grapple with many of the same challenges Hooke faced: resolution, contrast, field of view, optical aberrations, and obscurity of thick samples. (To overcome the thick sample problem, Hooke used a very sharp knife to cut an "exceeding thin" slice of cork &mdash; a technique [http://www.wired.com/wiredscience/2014/01/hm-brain-closeup/ still in everyday use].)
* Correct images for nonuniform illumination
+
  
====Week 3====
+
Hooke spent countless hours hand drawing the breathtaking illustrations for ''Micrographia''.  A CCD camera in the image plane of your microscope will provide a huge advantage. You will be able to record micrographs nearly as spectacular as Hooke's in a fraction of a second and with far less skill. (As a young man, Hooke apprenticed as a painter. The guy could draw.)
* Track microspheres suspended in a solvent
+
* Estimate diffusion coefficients; compute Avagadro's number
+
* Track the motion of vesicles under transport in a plant cell
+
* Estimate the number of myosin motors involved in vesicle transport
+
  
==Microscopy lab etiquette==
+
Specimens in part 1 will be illuminated by an LED that shines light through the sample plane. The illumination will show up as a bright background in your images. The unsurprising name of this method is: transilluminated, bright field microscopy. Transillumination works well for samples that absorb or scatter a lot of light. Most biological samples have low contrast when imaged this way. Despite the limitations of bright field microscopy, many important discoveries were made with this simple method. Hooke was an early discoverer of plant cells, but he was mostly interested in how the cell structure of his cork sample explained the material's unique mechanical properties. He soon trained his microscope on other things (like glass canes, a bloodsucking louse, and feathers).  
* Observe all laser safety guidelines.
+
* Keep all of the boxes for the optics you use with your instrument to simplify putting things away.  
+
* The stages are very expensive. To prevent accidents, ensure that there is a srew holding the post base to an optical breadboard or table at all times.
+
* There are not enough stages to go around. Remove the stage from your microscope and leave it at the lab station when you are done.
+
* Leave the illuminator at the lab station when you are done.
+
* Return objective lenses to the drawer when you are not using them.
+
* Never use an SM1T2 coupler without a locking ring &mdash; they are very difficult to remove if they are tightened against a lens tube or tube ring.
+
* Use tube rings (never an SM1T2) to mount optics in lens tubes.
+
  
==Readings==
+
[[Image:Barbara McClintock with Microscope.jpg|thumb|Barbara McClintock with her microscope]]
* From [http://www.microscopyu.com Nikon MicroscopyU]
+
Likely inspired by ''Micrographia'', a Dutch draper named Anton van Leeuwenhoek honed his lens-making skills and developed his own microscope. Van Leeuwenhoek was intensely interested in the tiny creatures he dubbed "animalcules" that he observed in water, blood, semen, and other specimens. Looking at samples of plaque from his own mouth, van Leeuwenhoek recorded: "I then most always saw, with great wonder, that in the said matter there were many very little living animalcules, very prettily a-moving. The biggest sort. . . had a very strong and swift motion, and shot through the water (or spittle) like a pike does through the water. Looking at the second sort. . . oft-times spun round like a top. . . and these were far more in number." (Sadly, the colorful term "animalcule" did not have as much staying power as "cell.") Van Leeuwenhoek discovered bacteria, protozoa, spermatozoa, rotifers, ''Hydra'', ''Volvox'', and parthenogenesis in aphids. He was truly the first microbiologist.
** [http://www.microscopyu.com/articles/formulas/formulasconjugate.html Conjugate planes in optical microscopy] Includes transmitted and reflected (epi) illumination.
+
** [http://www.microscopyu.com/articles/formulas/formulasri.html Snell's law]
+
** [http://www.microscopyu.com/articles/formulas/formulasresolution.html Resolution]
+
  
==Microscope design==
+
[[Image:20.309 130905 InstructorMicroscope1.png|thumb|20.309 microscope]]
Your microscope will be capable of two types of illumination: transmitted bright field and epi-fluorescence. It is probably easiest to design the trans illumination microscope first and then add a fluorescence capability.
+
Perhaps the most remarkable discovery ever made with nothing but a simple light microscope was genetic transposition. Barbara McClintock was a talented microscopist who developed a technique that enabled her to distinguish individual chromosomes in ''Zea mays'' (corn) plant cells. One important element of her method was that she prepared her samples by squashing them instead of cutting thin slices as Hooke did 300 years earlier. She observed genetic transposition through an optical microscope in 1944, nearly 10 years before the chemical structure of DNA was deciphered. Several decades elapsed before molecular techniques sufficiently sophisticated to confirm her discovery were developed.<ref>See, for example: McClintock, B. ''The origin and behavior of mutable loci in maize.'' PNAS. 1950; 36:344-355. [http://library.cshl.edu/archives/archives/bmcbio.htm], [http://library.cshl.edu/archives/archives/bmcres.htm], and Endersby, Jim. ''A Guinea Pig's History of Biology.'' Cambridge, Massachusetts: Harvard University Press; 2007.</ref> McClintock was awarded the Nobel Prize in Physiology or Medicine in 1983 for her discovery.
  
===Magnification===
+
An example microscope made by the instructors will be available in the lab for you to examine. Feel free to make improvements on this design. Mechanical stability will be crucial for the particle tracking experiments in parts 3 and 4 of the lab. The required stability specification will be achieved through good design and careful construction — not by indiscriminate over-tightening of screws.
A simple microscope consists of two lenses arranged in a 4-f configuration (as shown in the figure).  
+
  
[[Image:20.309Hw3Image4flens.JPG|center|thumb|400px|A 4-f lens system using lenses with focal lengths f1 and f2. The object and image distances (so and si, respectively) for the lenses are indicated.]]
+
===Part 2===
 +
[[Image:20.309-OnionMembranes.jpg|thumb|Fluorescence image from 20.309 microscope. Onion endothelial cell incubated with FM 4-64 dye (Invitrogen). <ref>See class stellar site for protocol. Oh & Yamaguchi, unpublished lab report</ref>]]
 +
The development of fluorescence microscopy has been the single most important rejoinder to the contrast problem (and more recently to the resolution problem). Fluorescence microscopes rely on special molecules in the sample called fluorophores that absorb photons of one wavelength and then turn around and emit photons of a longer wavelength. Optical filters separate excitation from emission, producing an image that shows only the cordial glow of the fluorescent molecules on a dark background. Filtering out the illumination provides much better contrast than transillumination. Excellent techniques exist for attaching fluorophores to molecules of interest. The three principal techniques are: fluorescent stains, immunofluorescence, and fluorescent proteins. Fluorescent stains such as DAPI are small molecules that bind to particular sites (the minor groove of DNA in the case of DAPI). Immunofluorescence exploits antibodies conjugated to fluorophores to label specific molecules. A [http://igene.lifetechnologies.com/isearch/antibody.do?parameters=attributenavigator:Conjugate%20Type:Alexa-Fluor-Dyes&icid=fr-alexa-2 dizzying array] of antibodies and dyes exists. Because they are amenable to genetic manipulations, fluorescent protein techniques have had perhaps the most profound impact on biological science. The 2008 Nobel Prize honored Osamu Shimomura, Martin Chalfie and Roger Tsien for developing the green fluorescent protein (GFP).  
  
 +
In part 2 of the lab, you will augment your microscope to support fluorescence imaging. To test the new capabilities of your microscope, you will image fluorescent microspheres and immunofluorescently labeled biological samples. You will use image processing techniques to correct the images for nonuniform illumination.
  
Bright field transmitted microscopy is the simplest and most common optical microscopy method. In this technique, photons from an illuminator pass through the sample, where the may be absorbed, diffracted, or refracted. (The sample us usually mounted on a glass slide.) An objective lens on the opposite side of the sample collects the light. Most modern objective lenses produce collimated light, which is focused by a tube lens to form an image.
+
===Part 3===
 +
In part 3, you will make quantitative measurements using fluorescence. You will image tiny, fluorescent microscpheres to measure the resolution of your microscope. The beads are so small they act essentially like point sources. You will also take movies of larger microspheres diffusing in solvents of different viscosities. You will use image processing and particle tracking techniques to measure the diffusion coefficient of the particles and estimate the viscosity of the solvents. The details of some of the solvents will be revealed to you and others will be unknown.
  
Illumination for epifluorescence microscopy reaches the sample through the objective lens &mdash; from the same side of the sample that is observed. Epi-fluorescence microscopy is normally used on samples that have been labled with a fluorescent molecule called a fluorophore. The (narrowband) illumination wavelength must match the absorption characteristic of the fluorophore. After becoming excited by a photon from the illuminator, fluorophores emit photons with a longer wavelength. A dichroic mirror in the microscope reflects the illumination wavelength but allows the emitted photons to pass through.
+
Then, you will use particle tracking to make quantitative measurements of a biological sample. Procedures vary from year to year. Details will be provided in class.
  
 +
The final report should consist of all 3 sections in a single file. In the final document, you may revise any part of the first two sections without penalty. Only the final report will be graded. You may not skip any of the intermediate reports.
  
 +
Follow the format suggested in the [[Microscopy report outline]].
  
 +
====Background materials and references====
 +
The following online materials provide useful background for this part of the microscopy lab.
  
An example microscope made by the instructors will be available in the lab for you to examine. Feel free to make improvements on this design. Stability will be crucial for the particle tracking experiments. The required specification will be achieved through good design and careful construction &mdash; not by mindlessly overtightening screws.
+
* [[Geometrical optics and ray tracing]]
 +
* [[Physical optics and resolution]]
 +
* [https://stellar.mit.edu/S/course/20/fa13/20.309/materials.html Lectures 1 through 9 of the 20.309 class]
 +
* From [http://www.microscopyu.com Nikon MicroscopyU]
 +
** [http://www.microscopyu.com/articles/formulas/formulasconjugate.html Conjugate planes in optical microscopy] (includes transmitted and reflected (epi) illumination)
 +
** [http://www.microscopyu.com/articles/formulas/formulasri.html Snell's law]
 +
** [http://www.microscopyu.com/articles/formulas/formulasresolution.html Resolution]
  
Some elements must be positioned precise distances apart; other distances are not critical. Use ray-tracing to determine when this is the case.
+
==Microscope design==
 +
[[Image:20.309 130909 4fdashed3.png|thumb|300px|4f microscope.]]
 +
The diagram on the right shows a microscope that consists of two, positive focal length lenses of focal lengths ''f<sub>1</sub>'' and ''f<sub>2</sub>'', with ''f<sub>1</sub>''<''f<sub>2</sub>''. The lens nearer to the object is called the objective lens. The lens closer to the image is called the tube lens &mdash; presumably because it resides inside the tube of an old fashioned microscope. The distance between the lenses is equal to the sum of their focal lengths, ''f<sub>1</sub>''+''f<sub>2</sub>''.  
  
===Bright field transmitted illumination imaging===
+
In this type of microscope, objects such as bloodsucking lice are placed a distance ''f<sub>1</sub>'' from the objective lens. The diagram shows rays emanating from two representative points on the object in blue and green. Optical ray tracing rules dictate that rays emerging from a single point in the focal plane of a lens are parallel or ''collimated'' after refraction. &ldquo;Collimate&rdquo; is a term frequently used in optics that means, &ldquo;to make parallel.&rdquo; Thus, all of the rays originating from a single point on the sample travel in parallel in the space between the two lenses.
Sketch out a rough design for your microscope on paper. Begin with the bright field illumination path.
+
  
Note:
+
The same ray tracing rule can be applied in reverse to ascertain what happens when a set of parallel rays strikes the tube lens. After refraction, parallel rays pass through a single point in the back focal plane of the tube lens. This forms an image at a distance of ''f<sub>2</sub>'' from the tube lens. You can verify by similar triangles that the magnification of the system is ''-f<sub>2</sub>''/''f<sub>1</sub>'', minus one times the ratio of the focal lengths of the lenses. (The sign is negative because the image is inverted.) The total length of the system from object to image is 2''f<sub>1</sub>''+2''f<sub>2</sub>''. Based on this observation, somebody decided that it would be clever to call this design a "4f" microscope, in spite of the fact that there are two different efs. It's no wonder that many people find engineers a bit curious. They are meticulous about some terms and remiss about others. Feel free to call it a 2''f<sub>1</sub>''+2''f<sub>2</sub>'' &lsquo;scope in your head.
* The Nikon objective lenses are designed to be paired with a 200 mm tube lens.
+
* Assume that the objectives behave as ideal plano-convex lenses.
+
* Fine focusing will be achieved by adjusting the height of the sample stage.
+
* Start the alignment with a 10× objective but progress to 40× and 100×.
+
* Use the LED illuminators for bright field transmitted light imaging.
+
* Put a quick connect in your design such that the camera CCD will end up 200mm from the back focal plane of the objective. Remember that the CCD is recessed inside the opening of the camers.
+
* The barrier filter has been installed in a lens tube already attached to the camera.
+
  
===Fluorescence imaging===
+
The microscope you build in this lab will be a 4f system, assuming you follow instructions reasonably well.  
Now sketch a layout that includes both the bright field and fluorescence illumination paths.
+
 
+
* Use a dichroic to direct the laser illumination toward the sample and to pass the emitted (fluorescent) light back through to the CCD.
+
* The barrier filter removes any light from the illumination laser that was not reflected by the dichroic.
+
* During fluorescent imaging, you will not use the bright field illuminator. Bright field capability is useful for first visualizing the sample and viewing what features are in the field of view. Your design should retain the ability to do both white light and fluorescent visualization.
+
* Verify your design with one of the lab instructors before beginning construction.
+
==Microscope components==
+
 
+
[[Image:Microscope Block Diagram.JPG|center|thumb|400px|Microscope Block Diagram]]
+
 
+
===Rigid optical construction===
+
The structure of your microscope will be built from a combination of cage and lens tube components from ThorLabs. (See the [http://www.thorlabs.com/navigation.cfm?Guide_ID=50 ThorLabs online catalog] for more details. Print catalogs are available in the lab.) Be sure you understand how to use cage cubes ([http://www.thorlabs.com/thorProduct.cfm?partNumber=C4W C4W]), cube optic mounts ([http://www.thorlabs.com/thorProduct.cfm?partNumber=B5C B5C]), and kinematic mounting plates ([http://www.thorlabs.com/thorProduct.cfm?partNumber=B4C B4C]). Please ask about any components you are not sure how to use.
+
 
+
An example microscope put together by the instructors will be available in the lab for you to look at. Feel free to make improvements on this design. Stability will be crucial for the particle tracking experiments. This will be achieved through good design and careful construction &mdash; not by mindlessly overtightening screws.
+
 
+
===Simple lenses===
+
Plano-convex spherical lenses are available with focal lengths of 25, 50, 75, 100, 125, and 200 mm. It is best to mount most optics in short (0.5" or 0.3") lens tubes. It is acceptable to mount a lens between the end of a tube and a tube ring or between two tube rings. In most cases, the convex side of the lens faces toward the collimated beam; the planar side goes toward the convergent rays. Plano-concave lenses with focal lengths of -30 and -50 are also available.
+
 
+
* Advice: Some students in the past have had difficulty with the ''Three of These Things'' game. Verify all optics before you use them by measuring the focal length with a ruler.
+
 
+
As you install lenses into your microscope, put a piece of tape on the lens tube showing focal length and orientation. This will help you both during constructino and put-away. Save the lens storage boxes and return components to the correct boxes when you are done.
+
 
+
Handle lenses only by the edges. If a lens is dirty, first remove grit with a blast of clean air or CO2. Clean the lens by wiping with a folded piece of lens paper wetted with a drop of methanol. (Do not touch the part of the tissue you use for cleaning with your fingers.) In some cases, it may be helpful to hold the folded lens tissue in a hemostat. Ask an instructor if you need help.
+
  
 
===Objective lenses===
 
===Objective lenses===
Please see the Nikon [http://www.microscopyu.com/articles/optics/objectiveintro.html Introduction to Microscope Objectives] at their excellent [http://www.microscopyu.com/index.html MicroscopyU] website.
+
[[Image:Microscope Objective Markings.png|thumb|300 px|right]]
 +
To make a microscope with high magnification and a lot of light gathering capacity, it is desirable to use an objective lens that has a large cross section and a short focal length. Unfortunately, simple lenses with these attributes exhibit terrible aberrations. (Hooke used a small ball of glass as his objective lens. It is a wonder that he was able to produce such detailed and accurate drawings from the distorted field of view he observed. It is likely that Hooke was a more patient person than you.) To reduce aberrations to an acceptable level, modern objective lenses consist of many optical elements in series.  
  
There are three objective lenses available in the lab: a 10×, a 40×, and a 100×. All of these are designed for a 200 mm tube
+
Despite the complexity (and added thickness) of all of the lenses inside modern objectives, the way they bend light is almost exactly the same as a single, simple lens with a [http://www.sabrizain.org/startrek/Astrometrics/Spatial_Flexure-rift.html spatial rift] in the middle. Yes, I mean the hoakey, ''Star Trek'' sort of spatial rift where one bit of space connects to another, discontiguous bit of space &mdash; perhaps resulting from a transporter malfunction during a tachyon burst that hit chroniton beam at the entrance of a wormhole.  
lens. An adapter ring converts the objective mounting threads to the SM1 threads used by the lens tube system.
+
  
* The back focal plane (BFP) of the objective coincides with the rear of the objective housing. This is equivalent to the focal plane of a simple lens.
+
As shown in the diagram, there is a special point on the optical axis in front of the objective that is analogous to the front focal point of a simple lens. Just like a simple lens, a point source at that location results in collimated rays propagating parallel to the optical axis out of the back end of the objective (yellow rays in the diagram). The manufacturer specifies the special point as a distance in millimeters in from the frontmost optic of the objective, called the ''working distance''. The working distance is printed on the side of most objectives after the letters &ldquo;WD.&rdquo;
* ''Working distance'' (WD) is the distance between the front end of the objective and the sample plane (when
+
the sample is in focus). Generally, the higher the magnification, the lower the working distance.
+
* The 100× objective is designed to be used with immersion oil, which provides an optical medium of pre-
+
determined refractive index (n = 1.5). When using the 100× objective, place a drop of oil on it. Bring the drop in con-
+
tact with the slide cover glass. After use, clean off excess oil by wicking it away with lens paper. Do not put samples away dirty. It is not necessary to use immersion oil for thin samples such as the Air Force Target or Ronchi Ruling.
+
  
===Sample stage===
+
On the other side of the objective, there is another point that corresponds to the back focus of a simple lens (green rays in the diagram). The location of that point varies as a function of the magnification of the objective. For high magnification objectives, the point is inside the barrel of the objective. It may be outside the objective for low power objectives. Regrettably, the manufacturer does not specify this distance.  
Precision [http://www.newport.com/562-Series-ULTRAlign-Precision-Multi-Axis-Positio/140089/1033/catalog.aspx Newport X/Y/Z stages] mounted on a post are available at each lab station. The stage setup is top-heavy. Avoid accidents by ensuring that the post base is always attached to an optical breadboard or table. Return the stage to the lab station when you are done with it.
+
  
The z-axis adjustment of the sample stage provides fine focusing.
+
Except for the spatial rift, it is nearly always safe to imagine that a complicated objective lens functions like a single, simple lens of a certain focal length. The analogous focal length is called the equivalent focal length, or EFL. You can find the EFL of an objective lens by the formula ''EFL''=<sup>''RTL''</sup>/<sub>''M''</sub>, where ''RTL'' is a mysterious quantity called the reference tube length and ''M'' is the magnification. M is printed on the side of the objective in a large font, usually followed by an &ldquo;x.&rdquo; RTL is a manufacturer-specific number that is not printed on the objective, or just about anywhere else. RTL is equal to the focal length of the tube lens embedded inside the microscope that the manufacturer wants to sell you. Since we are building our own microscope, we will have to figure out what sort of tube lens the manufacturer had in mind. The objectives you will use were made by Nikon, which means that they were probably designed for a 200 mm tube lens. Thus, for a Nikon objective, ''EFL''=<sup>200 mm</sup>/<sub>''M''</sub>. A 40x Nikon objective acts like an ''f''=5 mm lens (with a spatial rift of perhaps 30 mm).
  
===Fluorescence illumination===
+
The objective's numerical aperture, or NA, is printed on the objective after the magnification and a slash (40x/0.65, for example). NA is a measure of the light gathering capacity. More is better.
The fluorescent illumination source is a 5 mW, λ=532 nm green laser pointer. Do not begin working with the laser until you are familiar with laser safety procedures. If you missed laser safety training, please see an instructor.
+
  
*Wear the correct laser safety eyewear at all times &mdash; even when your laser is not on &mdash; since other groups may be using theirs.
+
The infinity symbol near the silver ring on the objective in the picture specifies the ''tube length.'' This is different than the RTL from before. The tube length is the distance in millimeters at which the objective is designed to form an image. Older objectives often have a finite tube length like 160 or 200. Lenses with a finite tube length are optimized for a slightly different optical design in which the sample plane is a bit farther away than the front focus and the objective forms a real image. An infinity symbol in this location indicates that the objective is designed to have the sample exactly in its front focal plane in order to produce collimated light. ''Infinity corrected'' objectives are intended to be used in a way that they do not form an image.
*Make sure the laser warning light outside the main door to the lab is flashing before you turn on a laser.
+
*Never point the laser toward other people.
+
*Laser light can emerge from the top of the objective lens. '''Never put your face directly above the objective.'''
+
  
===Beam expansion===
+
The number after the &infin;/ is the coverslip thickness. Aberrations in the objective are optimized for a specific type of coverslip. Some very fancy objectives have an adjustment ring that allows for different thicknesses.
Collimated light comes out of the laser pointer. The light should also be collimated when it reaches the sample (Kohler illumination). In order to provide a good image, the laser light should illuminate most of the field of view. The beam emerging from the laser pointer (abuot 1.1 mm) may not be wide enough. This means that the laser beam will need to be expanded. What expansion factor should you use for a 40× objective? The design of the beam expander is a tradeoff. Overexpansion will decrease the light intensity at the sample and may not give you enough power for imaging. Underexpansion will result in very uneven illumination.
+
  
===Dichroic mirror and barrier filter===
+
Other markings you'll find on objectives include an intricate yet arcane nomenclature and set of acronyms that describe its optical properties. [http://www.microscopyu.com/articles/optics/objectivespecs.html This excellent webpage] has a good explanation of the markings.  
The fluorophores we will emit light in the orange-red region of the visible spectrum (550-600 nm) when excited by the 532 nm green laser.
+
  
In any fluorescence system, a key concern is viewing only the emitted fluorescence photons, and eliminating any background light, especially from the illumination source. Two optical elements address the problem. A dichroic mirror passes light of one wavelength, and reflects light of another. The transmission spectrum for the 565DCXT dichroicis shown in Fig. 3. A barrier filter blocks a particular spectral region. We will use the E590LPv2 barrier filter.
+
[[Image:20.309_130813_SimpleMicroscopeDiagram.png|thumb|right|450px|Simple microscope with Nikon 10x objective, 200 mm tube lens, and CCD camera. The objective has an equivalent focal length of 20 mm and a working distance of 7 mm.]]
 +
A simple microscope made with a 10x Nikon objective is shown on the right. The working distance of the objective is 7 mm. As per the manufacturer's specification, an ''f''=200 mm tube lens is placed 200 mm from the objective. The tube lens forms an image at a distance of 200 mm.
  
Figure 1 shows the tranmittance of the dichroic mirror and barrier filter over a range of wavelengths. The barrier filter is essential for high-sensitivity fluorescence imaging. It will pass the red light from the LED in the illuminator for bright field imaging as well as the fluorescence photons.
+
To record images, a CCD camera is placed in the image plane. The CCD camera is essentially a grid of light detectors connected to a computer. Each detector or ''pixel'' measures 7.4x7.4 microns. The camera you will use has a rectangular array of 656x492 of pixels.
  
[[Image:20.309Hw3Imagespectra.JPG|center|thumb|400px|Figure 3: The transmission spectra for the 565DCXT dichroic and the E590LPv2 barrier filter]]
+
====Lenses====
 +
Plano-convex spherical lenses are available with focal lengths of 25, 50, 75, 100, 125, 150, 175, 200, and 250 mm. Plano-concave lenses with focal lengths of -30, -50, and -75 are also available. It is acceptable to mount a lens between the end of a tube and a tube ring or between two tube rings. To facilitate easy installation and removal, mount lenses in short (''e.g.'' 0.5") lens tubes. In most cases, the convex side of the lens faces toward the collimated beam. The flat side goes toward the convergent rays.
  
===CCD camera===
+
===Microscope design===
The microscope you will build does not have an eyepiece for direct visual observation. Instead, images will be captured with a Firewire-enabled CCD camera ([http://www.theimagingsource.com/en/products/cameras/firewire_mono/dmk21f04/overview/ DMK 21F04] from [http://www.theimagingsource.com/en/products/ The Imaging Source]). Its monochrome (black and white) sensor contains a grid of 640×480 square pixels that measure 5.6 μm on a side. An adapter ring converts the C-mount thread on the camera to SM1.
+
There are a few additional things to keep in mind as you design your microscope. In traditional microscopes, the objective is above the sample and the observer looks down through an eyepiece. This is not a good arrangement for many biological applications. Looking through the growth medium in a petri dish or well distorts the image, so you would prefer to observe from the same side the cells are growing on. If you turn the dish upside down, all of to goo runs out on to the microscope stage. To solve this, biological microscopes usually place the objective underneath the sample. (This also explains why the print on the objective is upside down.) The optical path is pretty long, almost half a meter. It would be inconvenient to lie on the floor while making observations or stand on a ladder while changing samples, so most inverted microscopes use a 45 degree mirror, like M1 in the diagram below, to redirect light sideways.
 +
[[Image:20.309 130911 YourMicroscope.png|center|thumb|400px|20.309 microscope block diagram]]
  
The barrier filter, SM1 adapter, and a quick-connect have been installed on the camera. Please leave the camera at the lab station when you are done with it. Camera software called ''IC Capture'' is installed on the labstation PCs.
+
[[Image:20.309 140204_Dichroic.png|thumb|150px|Dichroic mirror]]
 +
You are going to add fluorescence capability to your microscope in part 2. With a little bit of planning, you can build your part 1 &lsquo;scope in such a way that you won't have to take your instrument apart for part 2. One of the key components needed for epifluorescence is a dichroic mirror, DM in the diagram. Dichroic mirrors reflect some colors of light, but pass others. The one you will use reflects the green laser light used to excite the sample but passes red light emitted by fluorescent molecules in the sample.  
  
 +
[[Image:20.309Hw3Imagespectra.JPG|right|thumb|The transmission spectra for the 565DCXT dichroic and the E590LPv2 barrier filter]]
 +
The dichroic mirror is imperfect and still allows a substantial amount of green light to pass. It is thus supplemented by a barrier filter BF to attenuate green light by 5 orders of magnitude. The combination DM + BF is essential for making good images.
  
==Experiment 1:microscope construction and characterization==
+
In contrast with the transillumination bright-field approach where the LED incident light traverses the sample before reaching all image-forming optics, illumination for epi-fluorescence microscopy reaches the sample through the objective lens — from the same side of the sample that is observed. The rationale behind this design choice is the relative dimness of the fluorescence ''emission'' signal with respect to the needed ''excitation'' intensity.  Even though fluorescence excitation and emission occur at distinct wavelengths and can thus be filtered away from one another, it is still advantageous to restrict to the utmost the amount of excitation light in the image plane.
The microscope should be built in two stages. First, build a white-light inverted microscope. Verify its alignment and magnification.  
+
  
Next, add the fluorescence branch. An adjustable iris aperture and piece of lens or tissue paper can be very helpful for aligning the laser and directing it along the axis of the tubes.
 
  
===Bright field calibration===
+
In epi-fluorescence mode, the sample is illuminated by a green laser.  Traveling through and focused by the objective lens, this collimated laser beam would result in single point illumination of the sample.  To restore collimated excitation of a broader region of the sample, lens L5 must be inserted in the optical path... with the drawback that the laser beam's diameter, ~ 1.1 mm originally, thus becomes "minified" by a factor ''f<sub>5</sub> / f<sub>obj</sub>'', which would result in the illumination of a disc only ~ 25 &mu;m in diameter!
Make images the following samples using the three different objectives (10×, 40×, and 100×):
+
A Gallilean beam expander (L3 and L4) is thus the final requirement of the microscope design. It allows the collimated laser illumination to match the CCD camera field of view.  The focal lengths chosen for L3 and L4, and thus the expansion factor of the beam expander, are a tradeoff between uniformity and brightness of illumination, given the Gaussian shape of the laser beam.
  
* The smallest line pair on the Air Force imaging target
+
{{:Optical microscopy lab wiki pages}}
* A slide of 4 μm latex spheres
+
* Ronchi ruling - a periodic pattern containing 600 line-pairs per mm
+
  
Can you see all these samples? What is the magnification of the microscope and the size of its field of view? Is it what you expected?
+
==References==
  
==Experiment 2:Fluorescence characterization and imaging==
 
You have the following samples available for imaging using both 40× and 100× objectives:
 
 
*A fluorescence reference slide
 
*A sample slide with 4 μm red-fluorescent beads (modified with Nile Red dye, peak excitation at 535nm, peak emission at 575nm).
 
 
Imaging tasks:
 
 
#Use a fluorescence reference slide to center the field of view and to optimize the uniformity of illumination. Take an image of this. Use the histogram display of the camera software to be certain that the image is not overexposed.
 
#Take an image of the fluorescent beads. Perform flat field correction on the image(i.e. divide the image by the normalized reference image). Compare what you see before and after flat field correction.
 
#Image a stained onion slide in bright field and fluorescent contrast. Do a flat-field correction on the fluorescent image. Overlay the fluorescent image on top of the bright field image.
 
#Make an image of the PSF slide. This slide consists of 190nm fluorescent beads plus some 1&mu; beads to help focus in bright field contrast. You may want to make multiple images of the slide and average them to reduce noise. If you do so, be sure not to bump the table or make any adjustments in-between images.
 
 
==Experiment 3: particle tracking==
 
====Introduction and background<ref name="BMC">taken from http://labs.physics.berkeley.edu/mediawiki/index.php/Brownian_Motion_in_Cells</ref>====
 
If you have ever looked at an aqueous sample through a microscope, you have probably noticed that every small particle you see wiggles about continuously. Robert Brown, a British botanist, was not the first person to observe these motions, but perhaps the first person to recognize the significance of this observation. Experiments quickly established the basic features of these movements. Among other things, the magnitude of the fluctuations depended on the size of the particle, and there was no difference between "live" objects, such as plant pollen, and things such as rock dust. Apparently, finely crushed pieces of an Egyptian mummy also displayed these fluctuations.
 
 
Brown noted: ''[The movements] arose neither from currents in the fluid, nor from its gradual evaporation, but belonged to the particle itself''.
 
 
This effect may have remained a curiosity had it not been for A. Einstein and M. Smoluchowski. They realized that these particle movements made perfect sense in the context of the then developing kinetic theory of fluids. If matter is composed of atoms that collide frequently with other atoms, they reasoned, then even relatively large objects such as pollen grains would exhibit random movements. This last sentence contains the ingredients for several Nobel prizes!
 
 
Indeed, Einstein's interpretation of Brownian motion as the outcome of continuous bombardment by atoms immediately suggested a direct test of the atomic theory of matter. J. Perrin received the 1926 Nobel Prize for validating Einstein's predictions, thus confirming the atomic theory of matter.
 
 
Since then, the field has exploded, and a thorough understanding of Brownian motion is essential for everything from polymer physics to biophysics, aerodynamics, and statistical mechanics. One of the aims of this lab is to directly reproduce the experiments of J. Perrin that lead to his Nobel Prize. A translation of the key work is included in the reprints folder. Have a look – he used latex spheres, and we will use polystyrene spheres, but otherwise the experiments will be identical. In addition to reproducing Perrin's results, you will probe further by looking at the effect of varying solvent molecule size.
 
 
===Microscope stability for particle tracking===
 
To verify that your system is sufficiently stable for accurate particle tracking, measure a dry specimen containing 1&mu;  beads in bright field contrast. Chose a field of view in which you can see at least 3-4 beads. Using a 40x objective, track the beads for about 3 min. Use the bead tracking processing algorithm on two beads to calculate two trajectories. To further reduce common-mode motion from vibrations, calculate the differential trajectory from the individual trajectories of these two beads. Calculate the MSD from the differential trajectory. Your MSD should start out less than 10 nm<sup>2</sup> at t = 1 sec. and still be less than 100 nm2 for t = 180 sec.
 
 
===Estimating the diffusion coefficient by tracking suspended microspheres===
 
According to theory,<ref>A. Einstein, [http://www.math.princeton.edu/~mcmillen/molbio/papers/Einstein_diffusion1905.pdf On the Motion of Small Particles Suspended in Liquids at Rest Required by the Molecular-Kinetic Theory of Heat], Annalen der Physik (1905).</ref><ref>E. Frey and K. Kroy, [http://www3.interscience.wiley.com/cgi-bin/abstract/109884431/ Brownian motion: a paradigm of soft matter and biological physics], Ann. Phys. (2005). Published on the 100th anniversary of Einstein’s paper, this reference chronicles the history of Brownian motion from 1905 to the present.</ref><ref>R. Newburgh, [http://scitation.aip.org/journals/doc/AJPIAS-ft/vol_74/iss_6/478_1.html Einstein, Perrin, and the reality of atoms: 1905 revisited], Am. J. Phys. (2006). A modern replication of Perrin's experiment. Has a good, concise appendix with both the Einstein and Langevin derivations.</ref><ref>M. Haw, [http://stacks.iop.org/JPhysCM/14/7769 Colloidal suspensions, Brownian motion, molecular reality: a short history], J. Phys. Condens. Matter (2002).</ref> the mean squared displacement of a suspended particle is proportional to the time interval as: <math>\left \langle {\left | \vec r(t+\tau)-\vec r(t) \right \vert}^2 \right \rangle=2Dd\tau</math>, where <i>r</i>(<i>t</i>) = position, <i>d</i> = number of dimensions, <i>D</i> = diffusion coefficient, and <math>\tau</math>= time interval.
 
 
*Track some 1&mu; and 2&mu; microspheres and estimate the diffusion coefficient.
 
*Consider how many particles you should track and for how long. What is the uncertainty in your estimate?
 
 
See: [http://labs.physics.berkeley.edu/mediawiki/index.php/Simulating_Brownian_Motion this page] for more discussion of Brownian motion and a Matlab simulation.
 
 
==Experiment 3: Transport in a plant cell<ref name="BMC" />==
 
{|align="left"
 
| [[Image:BMC_Cytoskeleton.jpg|thumb|250px|center|The eukaryotic cytoskeleton. Actin filaments are shown in red, microtubules in green, and the nuclei are in blue.]]
 
|}
 
 
The third part of this experiment consists of observing the motion of particles inside a living cell. Cells transport food, waste, information, etc. in membrane-bound vesicles, which are visible under a light microscope. An old-fashioned view of a cell was that it is a "bag of water" containing various enzymes in which matter is transported passively by diffusion. Though diffusion is an important mechanism, it is too slow and random for long distance transport and directing materials where they are most needed, especially in larger cells.  It is now understood that cells have highly developed and intricate mechanisms for directed transport of materials. 
 
 
Most motions within and of cells involve two components, a cytoskeletal fiber that serves as a track, and a [http://en.wikipedia.org/wiki/Motor_protein motor protein] that does the work.  The motor molecule uses energy from the hydrolysis of one ATP molecule to bind to the fiber, bend to pull itself along the fiber, and release, all of which constitutes one "step".  For an animation of this stepping process, see this [http://valelab.ucsf.edu/images/movies/mov-procmotconvkinrev5.mov movie animation] from the Vale lab web site at UC San Francisco.  One can divide cellular motility mechanisms into  two classes based on the cytoskeletal fibers involved.  Microtubule-based mechanisms involve dynein or kinesin motors pulling on microtubules made of the protein tubulin. Actin-based mechanisms involve myosin motors pulling on actin fibers, also called microfibers. 
 
 
{|  align="center"
 
| [[Image:Myosin.gif|thumb|309px|Cartoon of myosin motors pulling organelles along an actin filament.]]|| [[Image:Kinesin.jpg|thumb|196px|Binding of kinesin  motor to microtubule.]]
 
|}
 
 
Virtually all cell types exhibit directed intracellular transport, but some cell types are particularly suitable for transport studies. Fish-scale pigment cells work well, since a large fraction of the cargoes that are transported are pigmented and thus easy to observe – the disadvantage is that you would need to bring a living fish into lab as a source of these cells. For convenience, we will use epidermal cells from onion bulbs that you can easily acquire in a grocery store.  With some care, a single layer of cells can be peeled off an inner section of the onion bulb and mounted flat on a slide. 
 
 
{|  align="center"
 
| [[Image:image005.png|thumb|250px|Onion cells in bright-field illumination. Round object in each cell is the nucleus.]] || [[Image:image003.png|thumb|250px|Vesicles in the cytoplasm of a plant cell, as seen in dark-field.]]
 
|}
 
 
In this experiment, we will be viewing the movement of vesicles within the cytoplasm of onion epidermal cells, shown above as they appear in bright-field and dark-field microscopy. The layers you see in an onion bulb develop into leaves when it sprouts.  Both sides of the leaf are covered with an epidermis consisting of brick-shaped cells, each with a cell wall and cell membrane on the outside.  Most of the interior of the cell is filled with a clear [http://en.wikipedia.org/wiki/Vacuole vacuole] that functions in storage and in maintenance of hydrostatic pressure essential to the stiffness of the plant (the difference between crisp lettuce and wilted lettuce). The [http://en.wikipedia.org/wiki/Cytoplasm cytoplasm], containing all of the other cell contents, occurs in a thin layer between the cell membrane and the vacuole, and in thin extensions through the vacuole called transvacuolar strands.  It is within the cytoplasm that you will be observing directed transport of vesicles by an actin-based mechanism.  These vesicles are spherical or rod-shaped organelles such as mitochondria, spherosomes, and [http://en.wikipedia.org/wiki/Peroxisome peroxisomes] ranging in size from 0.5 to 3 microns.  The diagram of an onion cell below shows the location of the cell wall, cytoplasm and vesicles in a typical cell; you will not be able to see much of the endoplasmic reticulum or the vacuole depicted because of their transparency.  Under the microscope, you will notice the vesicles are located just along the edges of the cell, or near the top and bottom surface if you focus up and down.  When  you see a narrow band of moving vesicles in the center of the cell, it is located in a transvacuolar strand, which may be a handy place to study motion.
 
 
[[Image:BMC_OnionStucture.gif|thumb|500 px|center|A 3D cross-section model of an onion epidermal cell, showing actin filaments and vesicles in the narrow bands of cytoplasm within the cell. ]]
 
 
In plant cells, vesicles are transported along actin fibers by myosin motor molecules.  An actin filament is composed of two intertwined actin chains, about 7 nm in diameter.  An actin fiber is considered structurally polar, having a (+) end and a (-) end, and most myosin motors move only towards the (+) end of the actin fiber.  In order to reverse the direction of a vesicle's motion, the vesicle must grab on to another actin fiber oriented in the opposite direction.  There are at least eighteen described classes of myosin, of which three, myosin VIII, XI, and XII are found in plant cells.  Some myosin motors are processive, meaning that they remain bound to an actin fiber as they move step-by-step along it (analagous to this [http://valelab.ucsf.edu/images/movies/mov-procmotconvkinrev5.mov movie animation of kinesin].  Other myosins are non-processive, releasing from the actin fiber after each step.  Myosin II found in muscle cells is non-processive, as illustrated in this [http://www.banyantree.org/jsale/actinmyosin/index.html video animation].  In the muscle functional unit, there are many myosin motors acting together, so there are always some attached to the actin fiber.  The myosin XI responsible for transport of plant cell vesicles is  the fastest myosin known and is processive.  It apparently is not known how many myosin molecules are attached to the surface of a vesicle or how many of those are active at one time in pulling the vesicle along an actin fiber. 
 
 
In some plant cells and algal cells, a large-scale streaming motion of the cytoplasm is observed, logically called [http://en.wikipedia.org/wiki/Cytoplasmic_streaming cytoplasmic streaming]. This bulk flow is believed to be caused by myosin motors pulling the extensive endoplasmic reticulum  along actin fibers lining the cell membrane.  Many other vesicles are then dragged along with the endoplasmic reticulum.  [http://www.ncbi.nlm.nih.gov/books/bv.fcgi?rid=mcb.figgrp.5242 Lodish and Berk, et al.] provide a detailed explanation of this process and a video of cytoplasmic streaming in the pond eeed Elodea can be viewed [http://www.microscopy-uk.org.uk/mag/imgnov00/cycloa3i.avi here].
 
 
In your observations of vesicles in onion epidermal cells, you should distinguish between the random Brownian motion of vesicles that are unattached (or at least not actively moving along) actin filaments, the directed transport of vesicles by attached myosin motors, and possibly (though we are not sure this really happens in onions) bulk flow of vesicles in cytoplasmic streaming.
 
 
 
 
 
 
==Report Requirements==
 
===Microscope documentation===
 
Make a sketch or block diagram of your full microscope setup (a hand drawing is perfectly acceptable, but please keep
 
it neat - a ruler is handy) with important parameters indicated (i.e. lens focal lengths, distances
 
of main components, etc.). It is not necessary to document the structural components.
 
 
===Bright field===
 
#Your favorite bright field images &mdash; one or two with each objective. Indicating the magnifcation and feld of view. Do they match what was expected?
 
#Calculate and include a table of the diffraction-limited resolution for the 10X, 40X and 100X objectives.
 
 
===Fluorescence===
 
#Filtered, normalized image you used for flat field correction with histogram and plot of one horizontal line near the middle.
 
#Uncorrected and flat-field corrected fluorescent image of 4&mu; beads.
 
#Image of PSF slide with 40X objective. Estimate the resolution using this objective, assuming the 190nm beads have negligible size. (Is this a good assumption? Can you improve on it?) One approach is to fit a Gaussian function to the image of one particle.
 
#Compare your estimate of the resolution of the 40X objective to the value you calculated.
 
 
===Particle tracking===
 
#Estimate the diffusion coefficient of the 1&mu; and 2&mu; microspheres using the particle tracks you recorded.
 
#Estimate the velocities of the granules being transported in the onion cell.
 
 
==References==
 
 
<References/>
 
<References/>
  
</div>
+
{{Template:20.309 bottom}}

Latest revision as of 15:19, 10 January 2017

20.309: Biological Instrumentation and Measurement

ImageBar 774.jpg

Hooke micrograph of cork cells

I took a good clear piece of Cork, and with a Pen-knife sharpen'd as keen as a Razor, I cut a piece of it off, and thereby left the surface of it exceeding smooth, then examining it very diligently with a Microscope, me thought I could perceive it to appear a little porous; but I could not so plainly distinguish them, as to be sure that they were pores, much less what Figure they were of: But judging from the lightness and yielding quality of the Cork, that certainly the texture could not be so curious, but that possibly, if I could use some further diligence, I might find it to be discernable with a Microscope, I with the same sharp Penknife, cut off from the former smooth surface an exceeding thin piece of it, and placing it on a black object Plate, because it was it self a white body, and casting the light on it with a deep plano-convex Glass, I could exceeding plainly perceive it to be all perforated and porous, much like a Honey-comb, but that the pores of it were not regular; yet it was not unlike a Honey-comb in these particulars.

I told several lines of these pores, and found that there were usually about threescore of these small Cells placed end-ways in the eighteenth part of an Inch in length, whence I concluded there must be neer eleven hundred of them, or somewhat more then a thousand in the length of an Inch, and therefore in a square Inch above a Million, or 1166400. and in a Cubick Inch, above twelve hundred Millions, or 1259712000. a thing almost incredible, did not our Microscope assure us of it by ocular demonstration.

Robert Hooke from Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon (1665)[1]


Introduction

In this lab, you will build an optical microscope using lenses, mirrors, filters, optical mounts, CCD cameras, lasers, and other components in the lab. The work is divided into 3 parts. Each part requires some lab work, some analysis, lots of clear thinking, and a written report. You will submit a short, group report after parts 1 and 2. The final report should include results from all 3 parts of the lab. You may revise and improve your part 1-2 reports before the final submission.

Part 1

Robert Hooke's microscope

In part 1 of the lab, you will build a compound microscope, determine its magnification, and attempt to measure the size of microscopic objects. The instrument you create will have a great deal in common with the microscope Robert Hooke built in the mid-1660s. Hooke meticulously documented his microscopic observations and published them in a popular volume called Micrographia in 1665. The measurements you make in part 1 will call to mind Hooke's early quantification of the size of plant cells (see quote at top of page). You will grapple with many of the same challenges Hooke faced: resolution, contrast, field of view, optical aberrations, and obscurity of thick samples. (To overcome the thick sample problem, Hooke used a very sharp knife to cut an "exceeding thin" slice of cork — a technique still in everyday use.)

Hooke spent countless hours hand drawing the breathtaking illustrations for Micrographia. A CCD camera in the image plane of your microscope will provide a huge advantage. You will be able to record micrographs nearly as spectacular as Hooke's in a fraction of a second and with far less skill. (As a young man, Hooke apprenticed as a painter. The guy could draw.)

Specimens in part 1 will be illuminated by an LED that shines light through the sample plane. The illumination will show up as a bright background in your images. The unsurprising name of this method is: transilluminated, bright field microscopy. Transillumination works well for samples that absorb or scatter a lot of light. Most biological samples have low contrast when imaged this way. Despite the limitations of bright field microscopy, many important discoveries were made with this simple method. Hooke was an early discoverer of plant cells, but he was mostly interested in how the cell structure of his cork sample explained the material's unique mechanical properties. He soon trained his microscope on other things (like glass canes, a bloodsucking louse, and feathers).

Barbara McClintock with her microscope

Likely inspired by Micrographia, a Dutch draper named Anton van Leeuwenhoek honed his lens-making skills and developed his own microscope. Van Leeuwenhoek was intensely interested in the tiny creatures he dubbed "animalcules" that he observed in water, blood, semen, and other specimens. Looking at samples of plaque from his own mouth, van Leeuwenhoek recorded: "I then most always saw, with great wonder, that in the said matter there were many very little living animalcules, very prettily a-moving. The biggest sort. . . had a very strong and swift motion, and shot through the water (or spittle) like a pike does through the water. Looking at the second sort. . . oft-times spun round like a top. . . and these were far more in number." (Sadly, the colorful term "animalcule" did not have as much staying power as "cell.") Van Leeuwenhoek discovered bacteria, protozoa, spermatozoa, rotifers, Hydra, Volvox, and parthenogenesis in aphids. He was truly the first microbiologist.

20.309 microscope

Perhaps the most remarkable discovery ever made with nothing but a simple light microscope was genetic transposition. Barbara McClintock was a talented microscopist who developed a technique that enabled her to distinguish individual chromosomes in Zea mays (corn) plant cells. One important element of her method was that she prepared her samples by squashing them instead of cutting thin slices as Hooke did 300 years earlier. She observed genetic transposition through an optical microscope in 1944, nearly 10 years before the chemical structure of DNA was deciphered. Several decades elapsed before molecular techniques sufficiently sophisticated to confirm her discovery were developed.[2] McClintock was awarded the Nobel Prize in Physiology or Medicine in 1983 for her discovery.

An example microscope made by the instructors will be available in the lab for you to examine. Feel free to make improvements on this design. Mechanical stability will be crucial for the particle tracking experiments in parts 3 and 4 of the lab. The required stability specification will be achieved through good design and careful construction — not by indiscriminate over-tightening of screws.

Part 2

Fluorescence image from 20.309 microscope. Onion endothelial cell incubated with FM 4-64 dye (Invitrogen). [3]

The development of fluorescence microscopy has been the single most important rejoinder to the contrast problem (and more recently to the resolution problem). Fluorescence microscopes rely on special molecules in the sample called fluorophores that absorb photons of one wavelength and then turn around and emit photons of a longer wavelength. Optical filters separate excitation from emission, producing an image that shows only the cordial glow of the fluorescent molecules on a dark background. Filtering out the illumination provides much better contrast than transillumination. Excellent techniques exist for attaching fluorophores to molecules of interest. The three principal techniques are: fluorescent stains, immunofluorescence, and fluorescent proteins. Fluorescent stains such as DAPI are small molecules that bind to particular sites (the minor groove of DNA in the case of DAPI). Immunofluorescence exploits antibodies conjugated to fluorophores to label specific molecules. A dizzying array of antibodies and dyes exists. Because they are amenable to genetic manipulations, fluorescent protein techniques have had perhaps the most profound impact on biological science. The 2008 Nobel Prize honored Osamu Shimomura, Martin Chalfie and Roger Tsien for developing the green fluorescent protein (GFP).

In part 2 of the lab, you will augment your microscope to support fluorescence imaging. To test the new capabilities of your microscope, you will image fluorescent microspheres and immunofluorescently labeled biological samples. You will use image processing techniques to correct the images for nonuniform illumination.

Part 3

In part 3, you will make quantitative measurements using fluorescence. You will image tiny, fluorescent microscpheres to measure the resolution of your microscope. The beads are so small they act essentially like point sources. You will also take movies of larger microspheres diffusing in solvents of different viscosities. You will use image processing and particle tracking techniques to measure the diffusion coefficient of the particles and estimate the viscosity of the solvents. The details of some of the solvents will be revealed to you and others will be unknown.

Then, you will use particle tracking to make quantitative measurements of a biological sample. Procedures vary from year to year. Details will be provided in class.

The final report should consist of all 3 sections in a single file. In the final document, you may revise any part of the first two sections without penalty. Only the final report will be graded. You may not skip any of the intermediate reports.

Follow the format suggested in the Microscopy report outline.

Background materials and references

The following online materials provide useful background for this part of the microscopy lab.

Microscope design

4f microscope.

The diagram on the right shows a microscope that consists of two, positive focal length lenses of focal lengths f1 and f2, with f1<f2. The lens nearer to the object is called the objective lens. The lens closer to the image is called the tube lens — presumably because it resides inside the tube of an old fashioned microscope. The distance between the lenses is equal to the sum of their focal lengths, f1+f2.

In this type of microscope, objects such as bloodsucking lice are placed a distance f1 from the objective lens. The diagram shows rays emanating from two representative points on the object in blue and green. Optical ray tracing rules dictate that rays emerging from a single point in the focal plane of a lens are parallel or collimated after refraction. “Collimate” is a term frequently used in optics that means, “to make parallel.” Thus, all of the rays originating from a single point on the sample travel in parallel in the space between the two lenses.

The same ray tracing rule can be applied in reverse to ascertain what happens when a set of parallel rays strikes the tube lens. After refraction, parallel rays pass through a single point in the back focal plane of the tube lens. This forms an image at a distance of f2 from the tube lens. You can verify by similar triangles that the magnification of the system is -f2/f1, minus one times the ratio of the focal lengths of the lenses. (The sign is negative because the image is inverted.) The total length of the system from object to image is 2f1+2f2. Based on this observation, somebody decided that it would be clever to call this design a "4f" microscope, in spite of the fact that there are two different efs. It's no wonder that many people find engineers a bit curious. They are meticulous about some terms and remiss about others. Feel free to call it a 2f1+2f2 ‘scope in your head.

The microscope you build in this lab will be a 4f system, assuming you follow instructions reasonably well.

Objective lenses

Microscope Objective Markings.png

To make a microscope with high magnification and a lot of light gathering capacity, it is desirable to use an objective lens that has a large cross section and a short focal length. Unfortunately, simple lenses with these attributes exhibit terrible aberrations. (Hooke used a small ball of glass as his objective lens. It is a wonder that he was able to produce such detailed and accurate drawings from the distorted field of view he observed. It is likely that Hooke was a more patient person than you.) To reduce aberrations to an acceptable level, modern objective lenses consist of many optical elements in series.

Despite the complexity (and added thickness) of all of the lenses inside modern objectives, the way they bend light is almost exactly the same as a single, simple lens with a spatial rift in the middle. Yes, I mean the hoakey, Star Trek sort of spatial rift where one bit of space connects to another, discontiguous bit of space — perhaps resulting from a transporter malfunction during a tachyon burst that hit chroniton beam at the entrance of a wormhole.

As shown in the diagram, there is a special point on the optical axis in front of the objective that is analogous to the front focal point of a simple lens. Just like a simple lens, a point source at that location results in collimated rays propagating parallel to the optical axis out of the back end of the objective (yellow rays in the diagram). The manufacturer specifies the special point as a distance in millimeters in from the frontmost optic of the objective, called the working distance. The working distance is printed on the side of most objectives after the letters “WD.”

On the other side of the objective, there is another point that corresponds to the back focus of a simple lens (green rays in the diagram). The location of that point varies as a function of the magnification of the objective. For high magnification objectives, the point is inside the barrel of the objective. It may be outside the objective for low power objectives. Regrettably, the manufacturer does not specify this distance.

Except for the spatial rift, it is nearly always safe to imagine that a complicated objective lens functions like a single, simple lens of a certain focal length. The analogous focal length is called the equivalent focal length, or EFL. You can find the EFL of an objective lens by the formula EFL=RTL/M, where RTL is a mysterious quantity called the reference tube length and M is the magnification. M is printed on the side of the objective in a large font, usually followed by an “x.” RTL is a manufacturer-specific number that is not printed on the objective, or just about anywhere else. RTL is equal to the focal length of the tube lens embedded inside the microscope that the manufacturer wants to sell you. Since we are building our own microscope, we will have to figure out what sort of tube lens the manufacturer had in mind. The objectives you will use were made by Nikon, which means that they were probably designed for a 200 mm tube lens. Thus, for a Nikon objective, EFL=200 mm/M. A 40x Nikon objective acts like an f=5 mm lens (with a spatial rift of perhaps 30 mm).

The objective's numerical aperture, or NA, is printed on the objective after the magnification and a slash (40x/0.65, for example). NA is a measure of the light gathering capacity. More is better.

The infinity symbol near the silver ring on the objective in the picture specifies the tube length. This is different than the RTL from before. The tube length is the distance in millimeters at which the objective is designed to form an image. Older objectives often have a finite tube length like 160 or 200. Lenses with a finite tube length are optimized for a slightly different optical design in which the sample plane is a bit farther away than the front focus and the objective forms a real image. An infinity symbol in this location indicates that the objective is designed to have the sample exactly in its front focal plane in order to produce collimated light. Infinity corrected objectives are intended to be used in a way that they do not form an image.

The number after the ∞/ is the coverslip thickness. Aberrations in the objective are optimized for a specific type of coverslip. Some very fancy objectives have an adjustment ring that allows for different thicknesses.

Other markings you'll find on objectives include an intricate yet arcane nomenclature and set of acronyms that describe its optical properties. This excellent webpage has a good explanation of the markings.

Simple microscope with Nikon 10x objective, 200 mm tube lens, and CCD camera. The objective has an equivalent focal length of 20 mm and a working distance of 7 mm.

A simple microscope made with a 10x Nikon objective is shown on the right. The working distance of the objective is 7 mm. As per the manufacturer's specification, an f=200 mm tube lens is placed 200 mm from the objective. The tube lens forms an image at a distance of 200 mm.

To record images, a CCD camera is placed in the image plane. The CCD camera is essentially a grid of light detectors connected to a computer. Each detector or pixel measures 7.4x7.4 microns. The camera you will use has a rectangular array of 656x492 of pixels.

Lenses

Plano-convex spherical lenses are available with focal lengths of 25, 50, 75, 100, 125, 150, 175, 200, and 250 mm. Plano-concave lenses with focal lengths of -30, -50, and -75 are also available. It is acceptable to mount a lens between the end of a tube and a tube ring or between two tube rings. To facilitate easy installation and removal, mount lenses in short (e.g. 0.5") lens tubes. In most cases, the convex side of the lens faces toward the collimated beam. The flat side goes toward the convergent rays.

Microscope design

There are a few additional things to keep in mind as you design your microscope. In traditional microscopes, the objective is above the sample and the observer looks down through an eyepiece. This is not a good arrangement for many biological applications. Looking through the growth medium in a petri dish or well distorts the image, so you would prefer to observe from the same side the cells are growing on. If you turn the dish upside down, all of to goo runs out on to the microscope stage. To solve this, biological microscopes usually place the objective underneath the sample. (This also explains why the print on the objective is upside down.) The optical path is pretty long, almost half a meter. It would be inconvenient to lie on the floor while making observations or stand on a ladder while changing samples, so most inverted microscopes use a 45 degree mirror, like M1 in the diagram below, to redirect light sideways.

20.309 microscope block diagram
Dichroic mirror

You are going to add fluorescence capability to your microscope in part 2. With a little bit of planning, you can build your part 1 ‘scope in such a way that you won't have to take your instrument apart for part 2. One of the key components needed for epifluorescence is a dichroic mirror, DM in the diagram. Dichroic mirrors reflect some colors of light, but pass others. The one you will use reflects the green laser light used to excite the sample but passes red light emitted by fluorescent molecules in the sample.

The transmission spectra for the 565DCXT dichroic and the E590LPv2 barrier filter

The dichroic mirror is imperfect and still allows a substantial amount of green light to pass. It is thus supplemented by a barrier filter BF to attenuate green light by 5 orders of magnitude. The combination DM + BF is essential for making good images.

In contrast with the transillumination bright-field approach where the LED incident light traverses the sample before reaching all image-forming optics, illumination for epi-fluorescence microscopy reaches the sample through the objective lens — from the same side of the sample that is observed. The rationale behind this design choice is the relative dimness of the fluorescence emission signal with respect to the needed excitation intensity. Even though fluorescence excitation and emission occur at distinct wavelengths and can thus be filtered away from one another, it is still advantageous to restrict to the utmost the amount of excitation light in the image plane.


In epi-fluorescence mode, the sample is illuminated by a green laser. Traveling through and focused by the objective lens, this collimated laser beam would result in single point illumination of the sample. To restore collimated excitation of a broader region of the sample, lens L5 must be inserted in the optical path... with the drawback that the laser beam's diameter, ~ 1.1 mm originally, thus becomes "minified" by a factor f5 / fobj, which would result in the illumination of a disc only ~ 25 μm in diameter! A Gallilean beam expander (L3 and L4) is thus the final requirement of the microscope design. It allows the collimated laser illumination to match the CCD camera field of view. The focal lengths chosen for L3 and L4, and thus the expansion factor of the beam expander, are a tradeoff between uniformity and brightness of illumination, given the Gaussian shape of the laser beam.

Optical microscopy lab

Code examples and simulations

Background reading

References

  1. Hooke, R. Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon London:Jo. Martyn, and Ja. Allestry, Printers to the Royal Society; 1665
  2. See, for example: McClintock, B. The origin and behavior of mutable loci in maize. PNAS. 1950; 36:344-355. [1], [2], and Endersby, Jim. A Guinea Pig's History of Biology. Cambridge, Massachusetts: Harvard University Press; 2007.
  3. See class stellar site for protocol. Oh & Yamaguchi, unpublished lab report