Difference between revisions of "Lab Manual: Optical Microscopy"

From Course Wiki
Jump to: navigation, search
(Rigid optical construction)
m (Introduction)
 
(356 intermediate revisions by 10 users not shown)
Line 1: Line 1:
 
[[Category:Lab Manuals]]
 
[[Category:Lab Manuals]]
 
[[Category:20.309]]
 
[[Category:20.309]]
<div style="padding: 10px; width: 7.5in; border: 3px solid #000000;">
+
[[Category:Optical Microscopy Lab]]
 +
{{Template:20.309}}
 +
[[Image:Hooke-CorkMicrograph.png|center|300px|Hooke micrograph of cork cells]]
  
==Introduction==
+
<blockquote>
In this lab, you will design and build a light microscope from optical components. Your instrument will be capable of two kinds of imaging: bright field transmitted light and fluorescence. In bright field,  
+
<div>
 +
''I took a good clear piece of Cork, and with a Pen-knife sharpen'd as keen as a Razor, I cut a piece of it off, and thereby left the surface of it exceeding smooth, then examining it very diligently with a Microscope, me thought I could perceive it to appear a little porous; but I could not so plainly distinguish them, as to be sure that they were pores, much less what Figure they were of: But judging from the lightness and yielding quality of the Cork, that certainly the texture could not be so curious, but that possibly, if I could use some further diligence, I might find it to be discernable with a Microscope, I with the same sharp Penknife, cut off from the former smooth surface an exceeding thin piece of it, and placing it on a black object Plate, because it was it self a white body, and casting the light on it with a deep plano-convex Glass, I could exceeding plainly perceive it to be all perforated and porous, much like a Honey-comb, but that the pores of it were not regular; yet it was not unlike a Honey-comb in these particulars.''
  
Bright field transmitted microscopy is perhaps the simplest and most common optical microscopy method. In this technique, photons from an illuminator pass through the sample, where the may be absorbed, diffracted, or refracted. (The sample us usually mounted on a glass slide.) An objective lens on the opposite side of the sample collects the light. Most modern objective lenses produce collimated light, which is focused by a tube lens to form an image.
+
''I told several lines of these pores, and found that there were usually about threescore of these small Cells placed end-ways in the eighteenth part of an Inch in length, whence I concluded there must be neer eleven hundred of them, or somewhat more then a thousand in the length of an Inch, and therefore in a square Inch above a Million, or 1166400. and in a Cubick Inch, above twelve hundred Millions, or 1259712000. a thing almost incredible, did not our Microscope assure us of it by ocular demonstration.''
  
Illumination for fluorescence microscopy normally reaches the sample through the objective lens &mdash; from the same side of the sample that is observed. Fluorescence microscopy is normally used on samples that have been labled with a fluorescent molecule called a fluorophore. The (narrowband) illumination wavelength must match the absorption characteristic of the fluorophore. After becoming excited by a photon from the illuminator, the fluorescent molecule will emit a photon of longer wavelength. A dichroic mirror in the microscope reflects the illumination wavelength but allows the emitted photons to pass through.
+
<blockquote>
 +
''&mdash; [http://en.wikipedia.org/wiki/Robert_Hooke Robert Hooke] from Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon (1665)<ref name="Micrographia">Hooke, R.  [http://www.gutenberg.org/files/15491/15491-h/15491-h.htm Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon] London:Jo. Martyn, and Ja. Allestry, Printers to the Royal Society; 1665</ref> ''
 +
</blockquote>
 +
</div>
 +
</blockquote>
 +
<br/>
  
The microscope design in thie lab is very flexible. Other contrast modes, such as dark field illumination and confocal, can be added to the microscope.
+
==Introduction==
 +
In this lab, you will build an optical microscope using lenses, mirrors, filters, optical mounts, CCD cameras, lasers, and other components in the lab. The work is divided into 3 parts. Each part requires some lab work, some analysis, lots of clear thinking, and a written report. You will submit a short, group report after parts 1 and 2. The final report should include results from all 3 parts of the lab. You may revise and improve your part 1-2 reports before the final submission.
  
==Objectives==
+
===Part 1===
* Learn about the theory and practice of light microscopy
+
[[Image:Hooke-Microscope.png|thumb|Robert Hooke's microscope]]
* Use ray tracing rules to design a transmitted bright field and fluorescent light microscope
+
In part 1 of the lab, you will build a compound microscope, determine its magnification, and attempt to measure the size of microscopic objects. The instrument you create will have a great deal in common with the microscope Robert Hooke built in the mid-1660s. Hooke meticulously documented his microscopic observations and published them in a popular volume called ''Micrographia'' in 1665. The measurements you make in part 1 will call to mind Hooke's early quantification of the size of plant cells (see quote at top of page). You will grapple with many of the same challenges Hooke faced: resolution, contrast, field of view, optical aberrations, and obscurity of thick samples. (To overcome the thick sample problem, Hooke used a very sharp knife to cut an "exceeding thin" slice of cork &mdash; a technique [http://www.wired.com/wiredscience/2014/01/hm-brain-closeup/ still in everyday use].)
* Construct the microscope from optical components
+
* Characterize the microscope's performance
+
* Enhance and analyze microscope images
+
* Quantitatively track moving particles in the field of view
+
  
==Roadmap and Milestones==
+
Hooke spent countless hours hand drawing the breathtaking illustrations for ''Micrographia''.  A CCD camera in the image plane of your microscope will provide a huge advantage. You will be able to record micrographs nearly as spectacular as Hooke's in a fraction of a second and with far less skill. (As a young man, Hooke apprenticed as a painter. The guy could draw.)
===Week 1===
+
# Design and Build a microscope
+
# Characterize the transmitted bright field performance of the microscope
+
# Add a laser illumination beam path
+
  
===Week 2===
+
Specimens in part 1 will be illuminated by an LED that shines light through the sample plane. The illumination will show up as a bright background in your images. The unsurprising name of this method is: transilluminated, bright field microscopy. Transillumination works well for samples that absorb or scatter a lot of light. Most biological samples have low contrast when imaged this way. Despite the limitations of bright field microscopy, many important discoveries were made with this simple method. Hooke was an early discoverer of plant cells, but he was mostly interested in how the cell structure of his cork sample explained the material's unique mechanical properties.  He soon trained his microscope on other things (like glass canes, a bloodsucking louse, and feathers).
# Characterize the fluorescent imaging performance of the microscope
+
# Make fluorescent images
+
# Correct for nonuniform illumination
+
# Track microspheres suspended in a solvent
+
# Estimated diffusion coefficients
+
# Track the motion of vesicles under transport in a plant cell
+
  
[[Image:20.309Hw3Imagelight.JPG|center|thumb|400px|Figure 1: Rough schematic showing white light and fluorescence light paths for imaging the specimen plane. Not drawn to scale.]]
+
[[Image:Barbara McClintock with Microscope.jpg|thumb|Barbara McClintock with her microscope]]
 +
Likely inspired by ''Micrographia'', a Dutch draper named Anton van Leeuwenhoek honed his lens-making skills and developed his own microscope. Van Leeuwenhoek was intensely interested in the tiny creatures he dubbed "animalcules" that he observed in water, blood, semen, and other specimens. Looking at samples of plaque from his own mouth, van Leeuwenhoek recorded: "I then most always saw, with great wonder, that in the said matter there were many very little living animalcules, very prettily a-moving. The biggest sort. . . had a very strong and swift motion, and shot through the water (or spittle) like a pike does through the water. Looking at the second sort. . . oft-times spun round like a top. . . and these were far more in number." (Sadly, the colorful term "animalcule" did not have as much staying power as "cell.") Van Leeuwenhoek discovered bacteria, protozoa, spermatozoa, rotifers, ''Hydra'', ''Volvox'', and parthenogenesis in aphids. He was truly the first microbiologist.  
  
==Microscope Construction==
+
[[Image:20.309 130905 InstructorMicroscope1.png|thumb|20.309 microscope]]
An example microscope made by the instructors will be available in the lab for you to look at. Feel free to make improvements on this design. Stability will be crucial for the particle tracking experiments. This will be achieved through good design and careful construction &mdash; not by mindlessly overtightening screws.
+
Perhaps the most remarkable discovery ever made with nothing but a simple light microscope was genetic transposition. Barbara McClintock was a talented microscopist who developed a technique that enabled her to distinguish individual chromosomes in ''Zea mays'' (corn) plant cells. One important element of her method was that she prepared her samples by squashing them instead of cutting thin slices as Hooke did 300 years earlier. She observed genetic transposition through an optical microscope in 1944, nearly 10 years before the chemical structure of DNA was deciphered. Several decades elapsed before molecular techniques sufficiently sophisticated to confirm her discovery were developed.<ref>See, for example: McClintock, B. ''The origin and behavior of mutable loci in maize.'' PNAS. 1950; 36:344-355. [http://library.cshl.edu/archives/archives/bmcbio.htm], [http://library.cshl.edu/archives/archives/bmcres.htm], and Endersby, Jim. ''A Guinea Pig's History of Biology.'' Cambridge, Massachusetts: Harvard University Press; 2007.</ref> McClintock was awarded the Nobel Prize in Physiology or Medicine in 1983 for her discovery.
  
===Microscopy lab etiquette===
+
An example microscope made by the instructors will be available in the lab for you to examine. Feel free to make improvements on this design. Mechanical stability will be crucial for the particle tracking experiments in parts 3 and 4 of the lab. The required stability specification will be achieved through good design and careful construction — not by indiscriminate over-tightening of screws.
* Observe all laser safety guidelines.
+
* Keep all of the boxes for the optics you use with your instrument to simplify putting things away.  
+
* The stages are very expensive. To prevent accidents, ensure that there is a srew holding the post base to an optical breadboard or table at all times.
+
* There are not enough stages to go around. Remove the stage from your microscope and leave it at the lab station when you are done.
+
* Leave the illuminator at the lab station when you are done.
+
* Return objective lenses to the drawer when you are not using them.
+
* Never use an SM1T2 coupler without a locking ring &mdash; they are very difficult to remove if they are tightened against a lens tube or tube ring.
+
* Use tube rings (never an SM1T2) to mount an optic in a lens tube.
+
  
===Microscope parts===
+
===Part 2===
 +
[[Image:20.309-OnionMembranes.jpg|thumb|Fluorescence image from 20.309 microscope. Onion endothelial cell incubated with FM 4-64 dye (Invitrogen). <ref>See class stellar site for protocol. Oh & Yamaguchi, unpublished lab report</ref>]]
 +
The development of fluorescence microscopy has been the single most important rejoinder to the contrast problem (and more recently to the resolution problem). Fluorescence microscopes rely on special molecules in the sample called fluorophores that absorb photons of one wavelength and then turn around and emit photons of a longer wavelength. Optical filters separate excitation from emission, producing an image that shows only the cordial glow of the fluorescent molecules on a dark background. Filtering out the illumination provides much better contrast than transillumination. Excellent techniques exist for attaching fluorophores to molecules of interest. The three principal techniques are: fluorescent stains, immunofluorescence, and fluorescent proteins. Fluorescent stains such as DAPI are small molecules that bind to particular sites (the minor groove of DNA in the case of DAPI). Immunofluorescence exploits antibodies conjugated to fluorophores to label specific molecules. A [http://igene.lifetechnologies.com/isearch/antibody.do?parameters=attributenavigator:Conjugate%20Type:Alexa-Fluor-Dyes&icid=fr-alexa-2 dizzying array] of antibodies and dyes exists. Because they are amenable to genetic manipulations, fluorescent protein techniques have had perhaps the most profound impact on biological science. The 2008 Nobel Prize honored Osamu Shimomura, Martin Chalfie and Roger Tsien for developing the green fluorescent protein (GFP).
  
====Rigid optical construction====
+
In part 2 of the lab, you will augment your microscope to support fluorescence imaging. To test the new capabilities of your microscope, you will image fluorescent microspheres and immunofluorescently labeled biological samples. You will use image processing techniques to correct the images for nonuniform illumination.
  
The design uses a combination to cage and lens tube components from ThorLabs. (See the [http://www.thorlabs.com/navigation.cfm?Guide_ID=50 ThorLabs online catalog] for more details. Print catalogs are available in the lab.) Be sure you understand how to use cage cubes ([http://www.thorlabs.com/thorProduct.cfm?partNumber=C4W C4W]), cube optic mounts ([http://www.thorlabs.com/thorProduct.cfm?partNumber=B5C B5C]), and kinematic mounting plates ([http://www.thorlabs.com/thorProduct.cfm?partNumber=B4C B4C]). Please ask about any components you are not sure how to use.
+
===Part 3===
 +
In part 3, you will make quantitative measurements using fluorescence. You will image tiny, fluorescent microscpheres to measure the resolution of your microscope. The beads are so small they act essentially like point sources. You will also take movies of larger microspheres diffusing in solvents of different viscosities. You will use image processing and particle tracking techniques to measure the diffusion coefficient of the particles and estimate the viscosity of the solvents. The details of some of the solvents will be revealed to you and others will be unknown.
  
====Simple lenses====
+
Then, you will use particle tracking to make quantitative measurements of a biological sample. Procedures vary from year to year. Details will be provided in class.
  
Plano-convex spherical lenses are available with focal lengths of 25, 50, 75, 100, 125, and 200 mm. It is best to mount most optics in short (0.5" or 0.3") lens tubes. It is acceptable to mount a lens between the end of a tube and a tube ring or between two tube rings. In most cases, the convex side of the lens faces toward the collimated beam; the planar side goes toward the convergent rays. Plano-concave lenses with focal lengths of -35 and -50 are also available.
+
The final report should consist of all 3 sections in a single file. In the final document, you may revise any part of the first two sections without penalty. Only the final report will be graded. You may not skip any of the intermediate reports.
  
* Advice: verify focal lengths of all optics. A few students in the past have had some difficulty with the ''Three of These Things'' game.
+
Follow the format suggested in the [[Microscopy report outline]].
  
As you install lenses into your microscope, put a piece of tape on the lens tube showing focal length and orientation. This will help you both during constructino and put-away. Save the lens storage boxes and return components to the correct boxes when you are done.
+
====Background materials and references====
 +
The following online materials provide useful background for this part of the microscopy lab.
  
Handle lenses only by the edges. If a lens is dirty, first remove grit with a blast of clean air or CO2. Clean the lens by wiping with a folded piece of lens paper wetted with a drop of methanol. (Do not touch the part of the tissue you use for cleaning with your fingers.) In some cases, it may be helpful to hold the folded lens tissue in a hemostat. Ask an instructor if you need help.
+
* [[Geometrical optics and ray tracing]]
 +
* [[Physical optics and resolution]]
 +
* [https://stellar.mit.edu/S/course/20/fa13/20.309/materials.html Lectures 1 through 9 of the 20.309 class]
 +
* From [http://www.microscopyu.com Nikon MicroscopyU]
 +
** [http://www.microscopyu.com/articles/formulas/formulasconjugate.html Conjugate planes in optical microscopy] (includes transmitted and reflected (epi) illumination)
 +
** [http://www.microscopyu.com/articles/formulas/formulasri.html Snell's law]
 +
** [http://www.microscopyu.com/articles/formulas/formulasresolution.html Resolution]
  
====Objective lenses====
+
==Microscope design==
Objectives are specialized lenses that are the workhorses of a microscope. They are designed with
+
[[Image:20.309 130909 4fdashed3.png|thumb|300px|4f microscope.]]
great effort (and often great expense) by the microscope manufacturers to have excellent optical
+
The diagram on the right shows a microscope that consists of two, positive focal length lenses of focal lengths ''f<sub>1</sub>'' and ''f<sub>2</sub>'', with ''f<sub>1</sub>''<''f<sub>2</sub>''. The lens nearer to the object is called the objective lens. The lens closer to the image is called the tube lens &mdash; presumably because it resides inside the tube of an old fashioned microscope. The distance between the lenses is equal to the sum of their focal lengths, ''f<sub>1</sub>''+''f<sub>2</sub>''.  
characteristics, low chromatic and geometric aberrations, and interchangeability. They are typically
+
swapped in and out to set the overall magnification - in a commercial microscope this is done
+
mechanically and sometimes automatically; we will do it manually.
+
  
For the 20.309 scopes, we have three objective lens options: a 10×, a 40×, and a 100×. They
+
In this type of microscope, objects such as bloodsucking lice are placed a distance ''f<sub>1</sub>'' from the objective lens. The diagram shows rays emanating from two representative points on the object in blue and green. Optical ray tracing rules dictate that rays emerging from a single point in the focal plane of a lens are parallel or ''collimated'' after refraction. &ldquo;Collimate&rdquo; is a term frequently used in optics that means, &ldquo;to make parallel.&rdquo; Thus, all of the rays originating from a single point on the sample travel in parallel in the space between the two lenses.
are referred to by their nominal magnification, which assumes they are used with a 200 mm tube
+
lens (which is how a commercial Nikon scope is built). We have adapter rings available that connect
+
the objective lens mounting threads to be SM1 lens tube system.
+
  
[[Image:20.309Hw3Imagenikon.JPG|right|thumb|300px|Figure 2: A typical Nikon objective lens]]
+
The same ray tracing rule can be applied in reverse to ascertain what happens when a set of parallel rays strikes the tube lens. After refraction, parallel rays pass through a single point in the back focal plane of the tube lens. This forms an image at a distance of ''f<sub>2</sub>'' from the tube lens. You can verify by similar triangles that the magnification of the system is ''-f<sub>2</sub>''/''f<sub>1</sub>'', minus one times the ratio of the focal lengths of the lenses. (The sign is negative because the image is inverted.) The total length of the system from object to image is 2''f<sub>1</sub>''+2''f<sub>2</sub>''. Based on this observation, somebody decided that it would be clever to call this design a "4f" microscope, in spite of the fact that there are two different efs. It's no wonder that many people find engineers a bit curious. They are meticulous about some terms and remiss about others. Feel free to call it a 2''f<sub>1</sub>''+2''f<sub>2</sub>'' &lsquo;scope in your head.
  
Three important things to note:
+
The microscope you build in this lab will be a 4f system, assuming you follow instructions reasonably well.  
*Objectives don't have an indicated focal distance; in-
+
stead, their back focal plane (BFP) is designed to coincide
+
with the rear of the objective housing. This is equivalent to
+
the focal location of a normal lens - ask a lab instructor
+
if you're unsure where this is.
+
*Another key parameter used to describe these lenses is
+
the ''working distance'' (WD). This is the distance between
+
the front end of the objective and the sample plane (when
+
the sample is in focus). Generally, the higher the magnification, the lower the working distance.
+
*The 100× objective is designed to be used with immersion oil, which provides an optical medium of pre-
+
determined refractive index (n = 1.5). When using the
+
100×, place a drop of oil on it, and bring the drop in con-
+
tact with the slide cover glass. After use, clean off excess oil by wicking it away with lens paper
+
(don't do a lot of rubbing). The 10× and 40× don't use oil.
+
  
====3.1.4 Sample holder/stage====
+
===Objective lenses===
The purpose of this component is to stably support the sample above the objective lens, and to
+
[[Image:Microscope Objective Markings.png|thumb|300 px|right]]
enable fine control for moving the sample in the x-y plane. Reliable translational control will be very
+
To make a microscope with high magnification and a lot of light gathering capacity, it is desirable to use an objective lens that has a large cross section and a short focal length. Unfortunately, simple lenses with these attributes exhibit terrible aberrations. (Hooke used a small ball of glass as his objective lens. It is a wonder that he was able to produce such detailed and accurate drawings from the distorted field of view he observed. It is likely that Hooke was a more patient person than you.) To reduce aberrations to an acceptable level, modern objective lenses consist of many optical elements in series.  
important especially at 40× or 100× magnification. Build the sample holder using a component
+
that allows x- and y-axis planar movement, and clamp it to a post.
+
  
====3.1.5 Fluorescence illumination====
+
Despite the complexity (and added thickness) of all of the lenses inside modern objectives, the way they bend light is almost exactly the same as a single, simple lens with a [http://www.sabrizain.org/startrek/Astrometrics/Spatial_Flexure-rift.html spatial rift] in the middle. Yes, I mean the hoakey, ''Star Trek'' sort of spatial rift where one bit of space connects to another, discontiguous bit of space &mdash; perhaps resulting from a transporter malfunction during a tachyon burst that hit chroniton beam at the entrance of a wormhole.  
We'll illuminate the fluorescent molecules of our scope using a λ=532 nm 5 mW green laser pointer
+
(see also Sections 3.1.6 and 3.2.2). DO NOT begin working with the laser in this section
+
until you're familiar with laser safety procedures! Use caution, because this laser can
+
damage your eyes.
+
  
 +
As shown in the diagram, there is a special point on the optical axis in front of the objective that is analogous to the front focal point of a simple lens. Just like a simple lens, a point source at that location results in collimated rays propagating parallel to the optical axis out of the back end of the objective (yellow rays in the diagram). The manufacturer specifies the special point as a distance in millimeters in from the frontmost optic of the objective, called the ''working distance''. The working distance is printed on the side of most objectives after the letters &ldquo;WD.&rdquo;
  
'''Laser Safety'''
+
On the other side of the objective, there is another point that corresponds to the back focus of a simple lens (green rays in the diagram). The location of that point varies as a function of the magnification of the objective. For high magnification objectives, the point is inside the barrel of the objective. It may be outside the objective for low power objectives. Regrettably, the manufacturer does not specify this distance.
  
Your lab instructor will provide a laser safety introduction before you start working.
+
Except for the spatial rift, it is nearly always safe to imagine that a complicated objective lens functions like a single, simple lens of a certain focal length. The analogous focal length is called the equivalent focal length, or EFL. You can find the EFL of an objective lens by the formula ''EFL''=<sup>''RTL''</sup>/<sub>''M''</sub>, where ''RTL'' is a mysterious quantity called the reference tube length and ''M'' is the magnification. M is printed on the side of the objective in a large font, usually followed by an &ldquo;x.&rdquo; RTL is a manufacturer-specific number that is not printed on the objective, or just about anywhere else. RTL is equal to the focal length of the tube lens embedded inside the microscope that the manufacturer wants to sell you. Since we are building our own microscope, we will have to figure out what sort of tube lens the manufacturer had in mind. The objectives you will use were made by Nikon, which means that they were probably designed for a 200 mm tube lens. Thus, for a Nikon objective, ''EFL''=<sup>200 mm</sup>/<sub>''M''</sub>. A 40x Nikon objective acts like an ''f''=5 mm lens (with a spatial rift of perhaps 30 mm).
Two major safety policies we need to follow for EHS compliance:
+
  
1. Wear the provided safety eyewear at all times, even when your laser is not on, since other
+
The objective's numerical aperture, or NA, is printed on the objective after the magnification and a slash (40x/0.65, for example). NA is a measure of the light gathering capacity. More is better.
groups may be running theirs.
+
  
2. Whenever you working with the lasers, turn on the laser warning light switch by the door to
+
The infinity symbol near the silver ring on the objective in the picture specifies the ''tube length.'' This is different than the RTL from before. The tube length is the distance in millimeters at which the objective is designed to form an image. Older objectives often have a finite tube length like 160 or 200. Lenses with a finite tube length are optimized for a slightly different optical design in which the sample plane is a bit farther away than the front focus and the objective forms a real image. An infinity symbol in this location indicates that the objective is designed to have the sample exactly in its front focal plane in order to produce collimated light. ''Infinity corrected'' objectives are intended to be used in a way that they do not form an image.  
the lab, and make sure the warning sign is flashing.
+
  
Additional common-sense practices to keep in mind:
+
The number after the &infin;/ is the coverslip thickness. Aberrations in the objective are optimized for a specific type of coverslip. Some very fancy objectives have an adjustment ring that allows for different thicknesses.
*Under no circumstances should you point the laser toward other people.
+
*In the configuration used here, laser light will emerge out of the top of the objective lens: do
+
not put your face directly above the objective.
+
  
 +
Other markings you'll find on objectives include an intricate yet arcane nomenclature and set of acronyms that describe its optical properties. [http://www.microscopyu.com/articles/optics/objectivespecs.html This excellent webpage] has a good explanation of the markings.
  
'''Beam expansion'''
+
[[Image:20.309_130813_SimpleMicroscopeDiagram.png|thumb|right|450px|Simple microscope with Nikon 10x objective, 200 mm tube lens, and CCD camera. The objective has an equivalent focal length of 20 mm and a working distance of 7 mm.]]
 +
A simple microscope made with a 10x Nikon objective is shown on the right. The working distance of the objective is 7 mm. As per the manufacturer's specification, an ''f''=200 mm tube lens is placed 200 mm from the objective. The tube lens forms an image at a distance of 200 mm.
  
The green light is collimated coming out of the laser pointer, and should also be collimated when it
+
To record images, a CCD camera is placed in the image plane. The CCD camera is essentially a grid of light detectors connected to a computer. Each detector or ''pixel'' measures 7.4x7.4 microns. The camera you will use has a rectangular array of 656x492 of pixels.
strikes the sample (this is called Kohler illumination). However, the green light should cover a good
+
portion of the field of view of your microscope, and the beam may not be wide enough emerging
+
from the pointer. This means that the laser beam will need to be expanded. What expansion factor
+
should you use for a 40× objective? You may want to be a bit conservative with beam expansion,
+
since the laser pointer is not too powerful and overexpansion will decrease the light intensity at the
+
sample and may not give you enough power for imaging.
+
Another simple 4-f system is useful here (in this instance the intermediate focal points must
+
coincide).
+
  
====3.1.6 Dichroic mirror and barrier filter====
+
====Lenses====
 +
Plano-convex spherical lenses are available with focal lengths of 25, 50, 75, 100, 125, 150, 175, 200, and 250 mm. Plano-concave lenses with focal lengths of -30, -50, and -75 are also available. It is acceptable to mount a lens between the end of a tube and a tube ring or between two tube rings. To facilitate easy installation and removal, mount lenses in short (''e.g.'' 0.5") lens tubes. In most cases, the convex side of the lens faces toward the collimated beam. The flat side goes toward the convergent rays.
  
The fluorophores we will fluoresce (emit) in the orange-red region of the visible spectrum (550-
+
===Microscope design===
600 nm). As mentioned, we'll excite these molecules using a 532 nm green laser pointer.
+
There are a few additional things to keep in mind as you design your microscope. In traditional microscopes, the objective is above the sample and the observer looks down through an eyepiece. This is not a good arrangement for many biological applications. Looking through the growth medium in a petri dish or well distorts the image, so you would prefer to observe from the same side the cells are growing on. If you turn the dish upside down, all of to goo runs out on to the microscope stage. To solve this, biological microscopes usually place the objective underneath the sample. (This also explains why the print on the objective is upside down.) The optical path is pretty long, almost half a meter. It would be inconvenient to lie on the floor while making observations or stand on a ladder while changing samples, so most inverted microscopes use a 45 degree mirror, like M1 in the diagram below, to redirect light sideways.
 +
[[Image:20.309 130911 YourMicroscope.png|center|thumb|400px|20.309 microscope block diagram]]
  
In any fluorescence system, a key concern is viewing only the emitted fluorescence photons, and
+
[[Image:20.309 140204_Dichroic.png|thumb|150px|Dichroic mirror]]
eliminating any background light, especially from the illumination source.
+
You are going to add fluorescence capability to your microscope in part 2. With a little bit of planning, you can build your part 1 &lsquo;scope in such a way that you won't have to take your instrument apart for part 2. One of the key components needed for epifluorescence is a dichroic mirror, DM in the diagram. Dichroic mirrors reflect some colors of light, but pass others. The one you will use reflects the green laser light used to excite the sample but passes red light emitted by fluorescent molecules in the sample.  
  
We have two pieces of optics to help us do this. Take a look at Figure 1 for the roles they
+
[[Image:20.309Hw3Imagespectra.JPG|right|thumb|The transmission spectra for the 565DCXT dichroic and the E590LPv2 barrier filter]]
play in the microscope layout. Note that the barrier filter is key for high-sensitivity fluorescence
+
The dichroic mirror is imperfect and still allows a substantial amount of green light to pass. It is thus supplemented by a barrier filter BF to attenuate green light by 5 orders of magnitude. The combination DM + BF is essential for making good images.
imaging, and will not significantly impact white-light imaging.
+
  
[[Image:20.309Hw3Imagespectra.JPG|center|thumb|400px|Figure 3: The transmission spectra for the 565DCXT dichroic and the E590LPv2 barrier filter]]
+
In contrast with the transillumination bright-field approach where the LED incident light traverses the sample before reaching all image-forming optics, illumination for epi-fluorescence microscopy reaches the sample through the objective lens — from the same side of the sample that is observed. The rationale behind this design choice is the relative dimness of the fluorescence ''emission'' signal with respect to the needed ''excitation'' intensity.  Even though fluorescence excitation and emission occur at distinct wavelengths and can thus be filtered away from one another, it is still advantageous to restrict to the utmost the amount of excitation light in the image plane.
  
  
'''dichroic mirror''' - Passes light of one wavelength, and reflects light of another. The transmission
+
In epi-fluorescence mode, the sample is illuminated by a green laser.  Traveling through and focused by the objective lens, this collimated laser beam would result in single point illumination of the sample.  To restore collimated excitation of a broader region of the sample, lens L5 must be inserted in the optical path... with the drawback that the laser beam's diameter, ~ 1.1 mm originally, thus becomes "minified" by a factor ''f<sub>5</sub> / f<sub>obj</sub>'', which would result in the illumination of a disc only ~ 25 &mu;m in diameter!
spectrum for our dichroics (the 565DCXT from Chroma Technology) is shown in Fig. 3.
+
A Gallilean beam expander (L3 and L4) is thus the final requirement of the microscope design. It allows the collimated laser illumination to match the CCD camera field of view.  The focal lengths chosen for L3 and L4, and thus the expansion factor of the beam expander, are a tradeoff between uniformity and brightness of illumination, given the Gaussian shape of the laser beam.
  
'''barrier filter''' - This filter blocks a particular spectral region. Our barrier filter is the E590LPv2,
+
{{:Optical microscopy lab wiki pages}}
also from Chroma Technology, and its spectrum is likewise shown below.
+
  
====3.1.7 CCD camera====
+
==References==
Unlike some microscopes you may be used to, the one we build will not have an
+
eyepiece for direct visual observation. Instead, we observe and capture images directly
+
with a Firewire-enabled CCD camera (DMK 21F04 from The Imaging Source). Its
+
monochrome (black and white) sensor is 640×480 pixels, each of which is a square
+
5.6 μm on a side. Like the objective lenses, we have an adapter ring to attach the
+
camera threads to the SM1 lens tube system.
+
  
The camera software is called IC Capture, and is run from the PC desktop. If the program
+
<References/>
gives an error and cannot find the connected camera, it may need to have its driver updated.
+
  
====3.2 Construction giudelines====
+
{{Template:20.309 bottom}}
A basic microscope is essentially a 4-f system (sketched in Fig. 4, which we have discussed in lecture.
+
The requirement of overlapping the focal points between the two lenses can sometimes be relaxed,
+
but sometimes needs to be adhered to very precisely. Do some ray-tracing to determine when this
+
is the case. Discuss with a lab instructor, if this is not clear.
+
 
+
[[Image:20.309Hw3Image4flens.JPG|center|thumb|400px|Figure 4: A 4-f lens system using lenses with focal lengths f1 and f2. The object and image distances (so and si, respectively) for the lenses are indicated.]]
+
 
+
====3.2.1 White light microscope====
+
The microscope should be built in two stages: first, you'll assemble a white-light inverted microscope, verify its alignment and magnification, and then you will add the fluorescence branch.
+
 
+
First sketch out a rough design for the microscope on paper. Verify this with one of the lab
+
instructors before beginning construction. Some hints and suggestions to help you with the layout
+
and design process:
+
 
+
* A rough schematic of the microscope geometry is shown in Figure 1.
+
* The Nikon objective lenses are designed to be paired with a 200 mm tube lens, which gives
+
the system the specified magnification.
+
* Assume that the objectives behave as ideal plano-convex lenses (this is what they are designed
+
to do). Since an objective's working distance can be quite short, the ability to finely control
+
its distance from the image is important. Mount your objective using a component that will
+
allow you to make fine distance adjustments.
+
* Start the alignment with a 10× objective but progress to 40× and 100×.
+
* Use the gooseneck lamps to trans-illuminate the sample for white light imaging.
+
* Use a quick-connect for the CCD camera. This will make modification much more convenient
+
for later construction stages.
+
 
+
====3.2.2 Fluorescence microscope====
+
As before, first sketch out a layout for adding fluorescence to the microscope and discuss it with
+
your lab instructor. Adding fluorescence should require only a few modifications to your setup.
+
 
+
Keep the following things in mind:
+
* Use the dichroic to direct the laser illumination toward the sample and to pass the emitted
+
(fluorescent) light back through to the CCD.
+
* An adjustable iris aperture and some lens or tissue paper can be very helpful for aligning the
+
laser and directing it along the axis of the tubes.
+
* The barrier filter is used to condition the image before it reaches the CCD camera, removing
+
any light from the illumination laser that might have gotten through the dichroic.
+
* During actual fluorescent imaging, you will not use white light illumination, but having white
+
light capability is useful for first visualizing the sample and viewing what features are in the
+
field of view. You should retain the ability to do both white light and fluorescent visualization.
+
 
+
==4 Experiment 1: Microscope Characterization and Fourier-Plane Imaging==
+
 
+
====4.1 White light calibration====
+
 
+
Use this white light microscope to image the following samples using the three different objectives
+
(10×, 40×, and 100×)
+
 
+
* The smallest line pair on the so-called Air Force imaging target test pattern.
+
 
+
* A slide of 4 μm latex spheres.
+
 
+
* Ronchi ruling - a periodic pattern containing 600 line-pairs per mm. (only use 40× and 100× objectives for this; why?)
+
 
+
Can you see all these samples? What is the magnification of the microscope and the size of its
+
field of view? Is it what you expected?
+
 
+
====4.2 Fluorescence characterization and imaging====
+
 
+
You have the following samples available for imaging using both 40× and 100× objectives:
+
 
+
(a) A solution slide of Rhodamine 6G (Rh6G) solution, used to adjust the microscope for uniform
+
illumination (this has its excitation peak at 530nm, and a fairly broad band of emission above
+
550nm).
+
 
+
(b) A sample slide with 4 μm red-fluorescent beads (modified with Nile Red dye, peak excitation
+
at 535nm, peak emission at 575nm).
+
 
+
Imaging tasks:
+
 
+
1. Use the Rh6G slide to optimize the uniformity of the illumination field (you want the light
+
distribution hitting the sample to be as uniform as possible). Take an image of this.
+
 
+
2. Measure the signal level while imaging Rh6G - try to achieve maximal uniformity and
+
brightness.
+
 
+
3. Image the red beads slide. Perform flat field correction on the beads (i.e. divide the bead
+
image by the normalized Rh6G image). Compare what you see before and after flat field correction.
+
 
+
''If the images are noticeably different, especially at 40× (i.e. there is significant non-uniformity in the illumination field), you should improve the scope's alignment until there isn't much difference.''
+
 
+
====4.3 Fourier optics====
+
Recall from lecture that a lens behaves as a Fourier transformer. When an object is placed at a
+
plane one focal length away from the lens, an image of the object's spatial frequencies is formed
+
one focal length away on the other side (called the Fourier plane). Likewise a Fourier transform
+
projected through a lens generates the corresponding image (see Figure 5).
+
 
+
====4.3.1 The light-scattering microscope====
+
Here we'll investigate the Fourier transforming properties of our microscope. This capability is
+
employed in a type of microscopy called Dynamic Light Scattering (DLS), which can be used, for
+
example, to quantify the presence of regular patterns or measure average sizes of features being
+
imaged.
+
 
+
To modify your microscope for imaging the Fourier plane, simply add a lens at the plane one
+
focal length behind the image plane and position the CCD camera one focal length behind the lens.
+
(A 50 mm lens is good { it will keep the beam path short, thought the exact focal length is not
+
critical.) Here, a pair of quick-connects will play a key role, letting you quickly switch between
+
capturing the image plane and the Fourier plane on the CCD.
+
 
+
{|
+
|[[Image:20.309Hw3Imagefourierplane.JPG|250px|right]]
+
|-align=center
+
|'''Figure 5: A lens system with unity magnification,
+
showing the relative locations of image and fourier
+
planes.'''
+
|}
+
To characterize the Fourier-imaging ability of
+
your microscope, we'll use the Ronchi ruling, illuminated by the laser. Its spatial frequency components can be readily imaged at the Fourier plane.
+
Before doing the experiment, estimate how
+
many diffraction orders you can observe with the
+
40£ and 100£ objectives (which have NA = 0:65
+
and NA = 1:3 respectively)?
+
The general equation describing the optical
+
geometry of a di®raction grating is
+
a(sin µm ¡ sin µi) = m¸ ;
+
where µi (the incident beam angle) and µm (the di®raction angle) are measured with respect to the
+
normal, a is the grating period, m is the di®raction order number, and ¸ is the light wavelength.
+
What is the incident angle of your experiment? What is the maximum di®raction angle allowed
+
by your apparatus? How is µm related to the numerical aperture of the objective lens? Recall that
+
NA = n sin µ, where n is the index of refraction of the medium. For air n = 1 and for oil n = 1:5.
+
Now, view the Fourier plane images of the Ronchi ruling with the 40£ and 100£ objective
+
lenses to verify your calculations.
+
 
+
{|
+
|[[Image:20.309Hw3Imagebutterfly.JPG|600px|left]]
+
|-align=center
+
|'''Figure 6: [Images and caption reproduced from Vukusic and Sambles] (a) Real color image of the blue
+
iridescence from a Morpho rhetenor wing. (b) Transmission electron micrograph (TEM) images showing
+
wing-scale cross-sections of M. rhetenor. (c) TEM images of a wing-scale cross-section of the related
+
species M. didius reveal its discretely con¯gured multilayer. The high occupancy and high layer number of
+
M. rhetenor in b creates an intense re°ectivity that contrasts with the more di®usely colored appearance
+
of M. didius, in which an overlying second layer of scales e®ects strong di®raction. Scale bars: (a) 1 cm, (b)
+
1.8 ¹m, (c) 1.3 ¹m. (d) Blue iridescence is prevalent in the fern-like tropical understory plants of the genus
+
Selaginella. (e) TEM section of a juvenile leaf from the plant Diplazium tomentosum.'''
+
|}
+
 
+
====4.3.2 Bio-photonic crystals====
+
 
+
Many biological systems in nature have interesting optical properties (please see the fascinating
+
review1 recently published in Nature). Some researchers have advocated using biological system
+
to \grow photonic crystals. Today, photonic crystals made by microlithography have been used
+
to guide light in optoelectronic systems and for optical computing. If photonic crystals can be
+
self-assembled using biological systems, this may become a new low-cost and high-throughput
+
manufacturing approach.
+
 
+
You are provided with two samples. (1) A peacock feather and (2) a piece of tissue paper.
+
Obtain the di®raction pattern from both samples using the 100£ objective. Now remove the lens
+
that generates the Fourier plane, and obtain a real image of both specimens. Can you explain the
+
di®raction patterns observed (quantitatively)? Do either of these samples exhibit the properties of
+
photonic crystals?
+
 
+
If this subject interests you, take a look at the additional references listed in the footnotes.2,3
+
 
+
 
+
==5 Experiment 2: Microrheology Measurements by Particle Tracking==
+
 
+
====5.1 Introduction and background====
+
 
+
Many cellular functions such as migration, differentiation, and proliferation are regulated by the
+
mechanical properties of cells, specifically, their elasticity and viscosity. Rheology is the science of
+
measuring materials' mechanical properties. Microrheology is a subgroup of techniques that are
+
capable of measuring mechanical property from microscopic material volumes. Clearly, given the
+
typical size of biological cells, microrheology is the technique needed to measure their elasticity and
+
viscosity.
+
 
+
The elastic and viscous properties of cells can be characterized by a complex-valued shear
+
modulus (with units of Pa) G*(w) = G'(w) + iG(w). The real part G'(w), referred to as the
+
storage modulus, is a measure of cell elasticity, while the imaginary part G(w), the loss modulus,
+
is a measure of their viscosity. A generalized Hookian relationship can be written as
+
 
+
F(w) math /propto /math G¤(!)¢r(!) ;
+
 
+
where ¢r(!) is a generalized displacement, and F(!) is a force linearly proportional to it via the
+
shear modulus. Therefore, we can measure the shear modulus if we can measure the deformation
+
of the cell under a known force. (Note that all these quantities are frequency-dependent).
+
Particle-tracking microrheometry is based on measuring the displacement of a particle with
+
radius a embedded in a cell driven by thermal forces (similarly to the vibrations of the cantilever that
+
you observed in the AFM lab). One complication is that this relationship is frequency dependent
+
{ this is because in complex °uids, such as the cellular cytoskeleton, there are di®erent energy
+
dissipation mechanisms over di®erent time scales.
+
 
+
To approach the derivation of the relevant formulas, it is more convenient to think in terms of
+
energy, rather than force. The relationship between stored energy and displacement has a familiar
+
form, similar to a spring-mass system (recall KE / k(¢z)2):
+
 
+
U(!) = Z F(!)dr / G¤(w)¢r2(!) :
+
 
+
What is the driving thermal energy U(!)? Recall also from that thermal energy is \white,
+
i.e., it contains equal power at all frequencies and is equal to 1
+
2kBT for each degree of freedom in a
+
second-order system, where kB is Boltzmann's constant and T is the absolute temperature. From
+
this relationship (since we're observing motion in two dimensions), we have
+
 
+
G¤(!) /
+
kBT
+
¢r2(!) :
+
 
+
Our argument is clearly very rough but a complete (and much more di±cult) derivation results
+
in the following equation (see Mason4 for details):
+
 
+
jG¤(!)j =
+
2kBT
+
3¼a h¢r2(!)i¡[1 + ®(!)]
+
 
+
Some key additional details to help you make sense of this equation:
+
1. As you can see, the dependence on displacement is more accurately expressed as the mean-
+
square displacement (MSD) h¢r2(!)i:
+
MSD = h¢r2(!)i = h¢r2(
+
+
¿
+
)i = h[r(t + ¿ ) ¡ r(t)]2i =
+
1
+
N
+
NX
+
i=1
+
[r(ti + ¿ ) ¡ r(ti)]2 ;
+
where h i denotes a time-average of the particle's displacement trajectory r(t), at discrete
+
times t = t1; : : : ; tn (as sampled by a digital system like the PC and camera). Additionally, ¿
+
is a characteristic lag/delay time for the measurement. corresponding to the frequency !.
+
2.
+
®(!) ´
+
@ lnh¢r2(¿ )i
+
@¿ ¯¯¯¯¯¿=2¼=!
+
:
+
3. The radius of the particle a plays a role in the formula.
+
4. ¡(¢) is the Gamma function (the generalized form of the factorial function, which can be
+
looked up in a mathematical table). Mason suggests that for our range of ®, ¡[1 + ®] ¼
+
¡0:457(1 + ®)2 ¡ 1:36(1 + ®) + 1:90:
+
This equation may look complicated but there is a simple approximation to calculate the elastic
+
and viscous moduli:
+
G0(!) = jG¤(!)j cos[¼®(!)=2]
+
G00(!) = jG¤(!)j sin[¼®(!)=2]
+
A detailed discussion of particle tracking microrheology can be found in the papers by Mason
+
and Lau5.
+
 
+
====5.2 Experimental details====
+
 
+
====5.2.1 Stability and setup====
+
 
+
The major challenge of particle tracking microrheometery is the small scale of the thermal forces and
+
the associated nanometer scale displacements. A few things you can do to ensure the experiment
+
works:
+
Be sure all the microscope components are rigidly assembled and ¯rmly tightened. Poorly built
+
scopes shake. It is also vital that when you perform this experiment that the optical tables are
+
°oating so that building noise is isolated. Of course, avoid touching the optical table and the
+
microscope during the measurement. There are also cardboard boxes available that you can put
+
over your microscope to isolate it from air currents. Finally, make sure that you and the people
+
around you are not talking too loudly during the experiment, because acoustic noise is signi¯cant.
+
The camera gain and brightness setting should be set as described in Figure 7.
+
 
+
====5.2.2 System veri¯cation====
+
To verify that your system is su±ciently rigid/stable, ¯rst measure a specimen containing 1 m red
+
°uorescent beads (Molecular Probes) dried in a cell dish. Chose a ¯eld of view in which you can
+
see at least 3-4 beads. Using a 40£ objective record an .avi movie for about 3 min. at a frame
+
rate of 30 frames/sec.
+
From your experience with image processing, you already know how to import .avi movie data
+
into matlab. To improve signal to noise ratio, sum every 30 frames together, which will make your
+
sequence have a temporal interval of 1 sec.
+
Use the bead tracking processing algorithm on two beads to calculate two trajectories. To
+
further reduce common-mode motion from room vibrations, calculate the di®erential trajectory
+
from the individual trajectories of these two beads. Calculate the MSD h¢r2i from this di®erential
+
trajectory. Your MSD should start out less than 10 nm2 at ¿ = 1 sec. and still be less than 100 nm2
+
for ¿ = 180 sec. If you don't get this, do not proceed further and ask for help.
+
5.2.3 Live cell measurements
+
Now that your system is su±ciently stable, you can run the experiment on cell samples. A key
+
technique to keep in mind when working with live cells { to avoid shocking them with \cold at
+
20±C, be sure that any solutions you add are pre-warmed to 37±C. We will keep a warm-water bath
+
running on the hotplate for this purpose, in which we will keep the various media.
+
You are provided with NIH 3T3 ibroblasts, which were prepared as follows:
+
Cells were cultured at 37±C in 5% CO2 in standard 100 mm £ 20 mm cell culture dishes
+
 
+
{|
+
|[[Image:20.309Hw3Imageic.JPG|600px|center]]
+
|-align=center
+
|'''Figure7: Proper camera gain settings for particle tracking: Choose Properties... under the Device menu and set the gain and brightness level to zero.  Change the exposure time until the intensity of the fluorescent beads is just high enough to achieve good contrast, but do not saturate the intensity value to 255'''
+
|}
+
 
+
(Corning) in a medium referred to as DMEM++ { this consists of DMEM (Cellgro) supplemented
+
with 10% fetal bovine serum (FBS - from Invitrogen) and 1% penicillin-streptomycin (Invitrogen).
+
The day prior to the microrheology experiments, ¯broblasts were plated on 35 mm glass-bottom
+
cell culture dishes (MatTek). On the day of the experiments, the cell con°uency should reach
+
about 60%. 1 ¹m diameter orange °uorescent microspheres (Molecular Probes) were mixed with
+
the growth medium (at a concentration of 5 £ 105 beads/mL) and added to the plated cells for a
+
period of 12 to 24 hours for bead endocytosis. Choose cells with 3 or 4 particles embedded in them
+
and take a movie as before. Take movies of about 3-5 cells.
+
Now treat the cell with the cytoskeleton-modifying chemical cytochalasin D (CytoD). Pipet
+
out the bu®er, add 1 mL CytoD solution at 10 ¹M (pre-mixed for you) to the dish, and wait for
+
20 min. It's a good idea to check on your cells after 20 min.: sometimes they are in bad shape
+
at that point but sometimes they still look very healthy. Wash and replace with bu®er twice with
+
2 mL pre-warmed DMEM++.
+
Repeat the particle tracking measurements again for 3-5 cells as quickly as you are able, since
+
their physiology has now been signi¯cantly disrupted and they will die within a couple of hours.
+
It's very unlikely that you'll be able to ¯nd the exact same cells you've already tracked; however
+
it's very much advisable to use the same dish for the \before and \after so you're aren't also
+
comparing between di®erent cell populations.
+
 
+
==6 Experiment 3: Fluorescence Imaging of the Actin Cytoskeleton==
+
 
+
We have now observed CytoD-induced rheological
+
changes of 3T3 ¯broblasts. This next experiment seeks
+
to better understand the e®ect of CytoD on cells. Since
+
there are signi¯cant rheological changes of 3T3 ¯brob-
+
last with CytoD, it is reasonable to assume that this
+
chemical may modify the ¯broblast cytoskeleton. The
+
most important component of the mammalian cell cy-
+
toskeleton is actin, so we will image actin structures in
+
these cells with and without CytoD treatment.
+
The actin cytoskeleton is visualized using phalloidin
+
labeled with Alexa Fluor 532. The excitation maximum
+
is at about 535nm and emission maximum is at about
+
575nm (see spectra in Fig. 8. Phalloidin is a fungal
+
toxin (small organic molecule) that binds only to poly-
+
merized ¯lamentous actin (F-actin), but not to actin
+
monomers, G-actin. (It is a toxin because it hinders
+
actin disassembly).
+
 
+
 
+
[[image]]
+
Figure 8: Excitation and emission spectra for
+
Alexa Fluor 532.
+
 
+
====6.1 Cell ¯xation and labeling protocol====
+
 
+
Prepare a dish of ¯xed ¯broblast cells (NIH 3T3) with actin labeled with phalloidin-Alexa Fluor
+
532. The labeling protocol is as follows:
+
The starting point is as before { cells cultured in dishes containing DMEM++, at approximately
+
60% con°uency. This is about the optimum percentage of cell population. If cells are too crowded,
+
they will not stretch properly and show their beautiful actin ¯laments.
+
Note also that these cells remain alive until the addition of formaldehyde, therefore requiring
+
that any bu®er/media added to be pre-warmed.
+
 
+
1. Pre-warm the 3.7% formaldehyde solution in a hot water bath on the hotplate.
+
2. Wash the cells twice with pre-warmed phosphate bu®ered saline (PBS) at pH 7.4.
+
{ Remove the medium with a pipette and wash each dish twice with 2 mL of PBS.
+
3. Fix the samples with 2 mL of 3.7% formaldehyde solution in PBS for 10 minutes at room
+
temperature.
+
{ \Fixing means cross-linking the intracellular proteins and freezing the cell structure. This
+
kills the cells.
+
4. Wash the cells three times with 2 mL PBS.
+
5. Extract each dish with 2 mL 0.1% Triton X-100 (a type of soap solution in PBS) for 3-5
+
minutes.
+
{ \Extraction refers to partially dissolving the plasma membrane of the cell.
+
6. Wash the cells 2-3 times with PBS.
+
7. Incubate the ¯xed cells with 2 mL 1% BSA (bovine serum albumin) in PBS for 20-30 minutes.
+
{ BSA blocks the nonspeci¯c binding sites.
+
8. Wash cells twice with PBS.
+
9. To each cell dish, add 200 ¹L of °uorescent phalloidin solution (speci¯c binding to F-actin)
+
pre-mixed in methanolic (diluted methanol). Carefully pipet this just onto the central circular
+
glass region of the dish, just enough to cover the cells, and incubate for 60 min. at room
+
temperature.
+
10. Wash three times with PBS.
+
11. You can now store the sample at +4±C (normal refrigerator) in PBS for a few days, or in
+
mounting medium for long-term storage (approx. up to 1 year).
+
By a very similar procedure, you should also prepare cells treated with CytoD. Between steps
+
2 and 3 of staining procedure, add 1 mL of the pre-warmed 10 ¹M CytoD solution for 20 min.
+
Afterwards, wash with DMEM++ twice.
+
 
+
====6.2 Actin imaging====
+
 
+
Since actin ¯laments and stress ¯bers are nm-scale objects, they are much dimmer than °uorescent
+
beads or the dye solution { care must be taken to get good images of the cytoskeleton. You may
+
need to cover the scope to reduce room light contamination.
+
Adjust the gain and the brightness of the camera to get the best picture. Be sure to keep the
+
same exposure conditions, however, for both untreated and treated cells.
+
Using the 40£ objective, take ¯ve images each of treated and untreated cells. You may have to
+
average multiple captured image frames to obtain acceptable signal to noise levels.
+
 
+
==7 Report Requirements==
+
 
+
''This lab report is due by 12:00 noon (in class) on Thursday, Nov. 30.''
+
 
+
====7.1 Microscope construction====
+
Make a sketch of your full microscope setup (a hand drawing is perfectly acceptable, but please keep
+
it neat - a ruler is handy) with important parameters indicated (i.e. lens focal lengths, distances
+
of main components, etc.).
+
 
+
====7.2 Experiment 1: microscope characterization and Fourier-plane imaging====
+
1. Include your favorite 2-3 white-light images, indicating the magni¯cation and ¯eld of view
+
(do they match what was expected?).
+
2. Include an uncorrected and a °at-¯eld-corrected °uorescent image of the 4 ¹m beads.
+
3. Calculate the di®raction-limited resolution for the 10£, 40£ and 100£ objectives, and how
+
this compares with which samples each objective could or could not resolve.
+
4. Include both the real and Fourier-plane images of the peacock feather and tissue paper, and
+
quantitatively describe how they relate to each other?
+
 
+
====7.3 Experiment 2: microrheology measurements by particle tracking====
+
Analyze the bead trajectories for the normal and CytoD-treated cells. Extract their MSD h¢r2i,
+
and the G0 and G00 modulus values, and include enough detail to make it clear how you performed
+
the analysis.
+
Do you get consistent results across multiple cells in each group? What can you tell about the
+
e®ect of CytoD on the mechanical properties of these cells?
+
 
+
====7.4 Experiment 3: °uorescence imaging of the actin cytoskeleton====
+
Include one or two of your favorite actin cytoskeleton images from each of the CytoD treated and
+
untreated cell groups. Use your image processing knowledge to optimize image quality.
+
Apply the algorithms you developed during the image processing lab to quantify the degree of
+
\¯berness of the actin structures with and without CytoD. As in Experiment 2, what can you say
+
about the e®ect of CytoD on these cells? Discuss whether the results of this experiment and the
+
microrheology experiment are consistent with each other?
+
 
+
 
+
====Bonus (optional)====
+
You've now used two di®erent optical microscopy methods to study the e®ects of cytochalasin D
+
on NIH 3T3 ¯broblasts. Hopefully, as you were working, many questions arose in your mind about
+
di®erent cell properties, their underlying physics, experimental conditions, etc.
+
For bonus credit, think about and propose any other experiments you might like to do (using
+
these microscopes, or the AFMs, or any other approach you like) to study any related questions of
+
cytoskeletal mechanics, microrheology, °uorescent labeling, etc. This is not a formal grant proposal
+
{ simply outline the question you'd like to answer and suggest an approach/method/technique that
+
you could use to test it.
+
 
+
</div>
+

Latest revision as of 15:19, 10 January 2017

20.309: Biological Instrumentation and Measurement

ImageBar 774.jpg

Hooke micrograph of cork cells

I took a good clear piece of Cork, and with a Pen-knife sharpen'd as keen as a Razor, I cut a piece of it off, and thereby left the surface of it exceeding smooth, then examining it very diligently with a Microscope, me thought I could perceive it to appear a little porous; but I could not so plainly distinguish them, as to be sure that they were pores, much less what Figure they were of: But judging from the lightness and yielding quality of the Cork, that certainly the texture could not be so curious, but that possibly, if I could use some further diligence, I might find it to be discernable with a Microscope, I with the same sharp Penknife, cut off from the former smooth surface an exceeding thin piece of it, and placing it on a black object Plate, because it was it self a white body, and casting the light on it with a deep plano-convex Glass, I could exceeding plainly perceive it to be all perforated and porous, much like a Honey-comb, but that the pores of it were not regular; yet it was not unlike a Honey-comb in these particulars.

I told several lines of these pores, and found that there were usually about threescore of these small Cells placed end-ways in the eighteenth part of an Inch in length, whence I concluded there must be neer eleven hundred of them, or somewhat more then a thousand in the length of an Inch, and therefore in a square Inch above a Million, or 1166400. and in a Cubick Inch, above twelve hundred Millions, or 1259712000. a thing almost incredible, did not our Microscope assure us of it by ocular demonstration.

Robert Hooke from Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon (1665)[1]


Introduction

In this lab, you will build an optical microscope using lenses, mirrors, filters, optical mounts, CCD cameras, lasers, and other components in the lab. The work is divided into 3 parts. Each part requires some lab work, some analysis, lots of clear thinking, and a written report. You will submit a short, group report after parts 1 and 2. The final report should include results from all 3 parts of the lab. You may revise and improve your part 1-2 reports before the final submission.

Part 1

Robert Hooke's microscope

In part 1 of the lab, you will build a compound microscope, determine its magnification, and attempt to measure the size of microscopic objects. The instrument you create will have a great deal in common with the microscope Robert Hooke built in the mid-1660s. Hooke meticulously documented his microscopic observations and published them in a popular volume called Micrographia in 1665. The measurements you make in part 1 will call to mind Hooke's early quantification of the size of plant cells (see quote at top of page). You will grapple with many of the same challenges Hooke faced: resolution, contrast, field of view, optical aberrations, and obscurity of thick samples. (To overcome the thick sample problem, Hooke used a very sharp knife to cut an "exceeding thin" slice of cork — a technique still in everyday use.)

Hooke spent countless hours hand drawing the breathtaking illustrations for Micrographia. A CCD camera in the image plane of your microscope will provide a huge advantage. You will be able to record micrographs nearly as spectacular as Hooke's in a fraction of a second and with far less skill. (As a young man, Hooke apprenticed as a painter. The guy could draw.)

Specimens in part 1 will be illuminated by an LED that shines light through the sample plane. The illumination will show up as a bright background in your images. The unsurprising name of this method is: transilluminated, bright field microscopy. Transillumination works well for samples that absorb or scatter a lot of light. Most biological samples have low contrast when imaged this way. Despite the limitations of bright field microscopy, many important discoveries were made with this simple method. Hooke was an early discoverer of plant cells, but he was mostly interested in how the cell structure of his cork sample explained the material's unique mechanical properties. He soon trained his microscope on other things (like glass canes, a bloodsucking louse, and feathers).

Barbara McClintock with her microscope

Likely inspired by Micrographia, a Dutch draper named Anton van Leeuwenhoek honed his lens-making skills and developed his own microscope. Van Leeuwenhoek was intensely interested in the tiny creatures he dubbed "animalcules" that he observed in water, blood, semen, and other specimens. Looking at samples of plaque from his own mouth, van Leeuwenhoek recorded: "I then most always saw, with great wonder, that in the said matter there were many very little living animalcules, very prettily a-moving. The biggest sort. . . had a very strong and swift motion, and shot through the water (or spittle) like a pike does through the water. Looking at the second sort. . . oft-times spun round like a top. . . and these were far more in number." (Sadly, the colorful term "animalcule" did not have as much staying power as "cell.") Van Leeuwenhoek discovered bacteria, protozoa, spermatozoa, rotifers, Hydra, Volvox, and parthenogenesis in aphids. He was truly the first microbiologist.

20.309 microscope

Perhaps the most remarkable discovery ever made with nothing but a simple light microscope was genetic transposition. Barbara McClintock was a talented microscopist who developed a technique that enabled her to distinguish individual chromosomes in Zea mays (corn) plant cells. One important element of her method was that she prepared her samples by squashing them instead of cutting thin slices as Hooke did 300 years earlier. She observed genetic transposition through an optical microscope in 1944, nearly 10 years before the chemical structure of DNA was deciphered. Several decades elapsed before molecular techniques sufficiently sophisticated to confirm her discovery were developed.[2] McClintock was awarded the Nobel Prize in Physiology or Medicine in 1983 for her discovery.

An example microscope made by the instructors will be available in the lab for you to examine. Feel free to make improvements on this design. Mechanical stability will be crucial for the particle tracking experiments in parts 3 and 4 of the lab. The required stability specification will be achieved through good design and careful construction — not by indiscriminate over-tightening of screws.

Part 2

Fluorescence image from 20.309 microscope. Onion endothelial cell incubated with FM 4-64 dye (Invitrogen). [3]

The development of fluorescence microscopy has been the single most important rejoinder to the contrast problem (and more recently to the resolution problem). Fluorescence microscopes rely on special molecules in the sample called fluorophores that absorb photons of one wavelength and then turn around and emit photons of a longer wavelength. Optical filters separate excitation from emission, producing an image that shows only the cordial glow of the fluorescent molecules on a dark background. Filtering out the illumination provides much better contrast than transillumination. Excellent techniques exist for attaching fluorophores to molecules of interest. The three principal techniques are: fluorescent stains, immunofluorescence, and fluorescent proteins. Fluorescent stains such as DAPI are small molecules that bind to particular sites (the minor groove of DNA in the case of DAPI). Immunofluorescence exploits antibodies conjugated to fluorophores to label specific molecules. A dizzying array of antibodies and dyes exists. Because they are amenable to genetic manipulations, fluorescent protein techniques have had perhaps the most profound impact on biological science. The 2008 Nobel Prize honored Osamu Shimomura, Martin Chalfie and Roger Tsien for developing the green fluorescent protein (GFP).

In part 2 of the lab, you will augment your microscope to support fluorescence imaging. To test the new capabilities of your microscope, you will image fluorescent microspheres and immunofluorescently labeled biological samples. You will use image processing techniques to correct the images for nonuniform illumination.

Part 3

In part 3, you will make quantitative measurements using fluorescence. You will image tiny, fluorescent microscpheres to measure the resolution of your microscope. The beads are so small they act essentially like point sources. You will also take movies of larger microspheres diffusing in solvents of different viscosities. You will use image processing and particle tracking techniques to measure the diffusion coefficient of the particles and estimate the viscosity of the solvents. The details of some of the solvents will be revealed to you and others will be unknown.

Then, you will use particle tracking to make quantitative measurements of a biological sample. Procedures vary from year to year. Details will be provided in class.

The final report should consist of all 3 sections in a single file. In the final document, you may revise any part of the first two sections without penalty. Only the final report will be graded. You may not skip any of the intermediate reports.

Follow the format suggested in the Microscopy report outline.

Background materials and references

The following online materials provide useful background for this part of the microscopy lab.

Microscope design

4f microscope.

The diagram on the right shows a microscope that consists of two, positive focal length lenses of focal lengths f1 and f2, with f1<f2. The lens nearer to the object is called the objective lens. The lens closer to the image is called the tube lens — presumably because it resides inside the tube of an old fashioned microscope. The distance between the lenses is equal to the sum of their focal lengths, f1+f2.

In this type of microscope, objects such as bloodsucking lice are placed a distance f1 from the objective lens. The diagram shows rays emanating from two representative points on the object in blue and green. Optical ray tracing rules dictate that rays emerging from a single point in the focal plane of a lens are parallel or collimated after refraction. “Collimate” is a term frequently used in optics that means, “to make parallel.” Thus, all of the rays originating from a single point on the sample travel in parallel in the space between the two lenses.

The same ray tracing rule can be applied in reverse to ascertain what happens when a set of parallel rays strikes the tube lens. After refraction, parallel rays pass through a single point in the back focal plane of the tube lens. This forms an image at a distance of f2 from the tube lens. You can verify by similar triangles that the magnification of the system is -f2/f1, minus one times the ratio of the focal lengths of the lenses. (The sign is negative because the image is inverted.) The total length of the system from object to image is 2f1+2f2. Based on this observation, somebody decided that it would be clever to call this design a "4f" microscope, in spite of the fact that there are two different efs. It's no wonder that many people find engineers a bit curious. They are meticulous about some terms and remiss about others. Feel free to call it a 2f1+2f2 ‘scope in your head.

The microscope you build in this lab will be a 4f system, assuming you follow instructions reasonably well.

Objective lenses

Microscope Objective Markings.png

To make a microscope with high magnification and a lot of light gathering capacity, it is desirable to use an objective lens that has a large cross section and a short focal length. Unfortunately, simple lenses with these attributes exhibit terrible aberrations. (Hooke used a small ball of glass as his objective lens. It is a wonder that he was able to produce such detailed and accurate drawings from the distorted field of view he observed. It is likely that Hooke was a more patient person than you.) To reduce aberrations to an acceptable level, modern objective lenses consist of many optical elements in series.

Despite the complexity (and added thickness) of all of the lenses inside modern objectives, the way they bend light is almost exactly the same as a single, simple lens with a spatial rift in the middle. Yes, I mean the hoakey, Star Trek sort of spatial rift where one bit of space connects to another, discontiguous bit of space — perhaps resulting from a transporter malfunction during a tachyon burst that hit chroniton beam at the entrance of a wormhole.

As shown in the diagram, there is a special point on the optical axis in front of the objective that is analogous to the front focal point of a simple lens. Just like a simple lens, a point source at that location results in collimated rays propagating parallel to the optical axis out of the back end of the objective (yellow rays in the diagram). The manufacturer specifies the special point as a distance in millimeters in from the frontmost optic of the objective, called the working distance. The working distance is printed on the side of most objectives after the letters “WD.”

On the other side of the objective, there is another point that corresponds to the back focus of a simple lens (green rays in the diagram). The location of that point varies as a function of the magnification of the objective. For high magnification objectives, the point is inside the barrel of the objective. It may be outside the objective for low power objectives. Regrettably, the manufacturer does not specify this distance.

Except for the spatial rift, it is nearly always safe to imagine that a complicated objective lens functions like a single, simple lens of a certain focal length. The analogous focal length is called the equivalent focal length, or EFL. You can find the EFL of an objective lens by the formula EFL=RTL/M, where RTL is a mysterious quantity called the reference tube length and M is the magnification. M is printed on the side of the objective in a large font, usually followed by an “x.” RTL is a manufacturer-specific number that is not printed on the objective, or just about anywhere else. RTL is equal to the focal length of the tube lens embedded inside the microscope that the manufacturer wants to sell you. Since we are building our own microscope, we will have to figure out what sort of tube lens the manufacturer had in mind. The objectives you will use were made by Nikon, which means that they were probably designed for a 200 mm tube lens. Thus, for a Nikon objective, EFL=200 mm/M. A 40x Nikon objective acts like an f=5 mm lens (with a spatial rift of perhaps 30 mm).

The objective's numerical aperture, or NA, is printed on the objective after the magnification and a slash (40x/0.65, for example). NA is a measure of the light gathering capacity. More is better.

The infinity symbol near the silver ring on the objective in the picture specifies the tube length. This is different than the RTL from before. The tube length is the distance in millimeters at which the objective is designed to form an image. Older objectives often have a finite tube length like 160 or 200. Lenses with a finite tube length are optimized for a slightly different optical design in which the sample plane is a bit farther away than the front focus and the objective forms a real image. An infinity symbol in this location indicates that the objective is designed to have the sample exactly in its front focal plane in order to produce collimated light. Infinity corrected objectives are intended to be used in a way that they do not form an image.

The number after the ∞/ is the coverslip thickness. Aberrations in the objective are optimized for a specific type of coverslip. Some very fancy objectives have an adjustment ring that allows for different thicknesses.

Other markings you'll find on objectives include an intricate yet arcane nomenclature and set of acronyms that describe its optical properties. This excellent webpage has a good explanation of the markings.

Simple microscope with Nikon 10x objective, 200 mm tube lens, and CCD camera. The objective has an equivalent focal length of 20 mm and a working distance of 7 mm.

A simple microscope made with a 10x Nikon objective is shown on the right. The working distance of the objective is 7 mm. As per the manufacturer's specification, an f=200 mm tube lens is placed 200 mm from the objective. The tube lens forms an image at a distance of 200 mm.

To record images, a CCD camera is placed in the image plane. The CCD camera is essentially a grid of light detectors connected to a computer. Each detector or pixel measures 7.4x7.4 microns. The camera you will use has a rectangular array of 656x492 of pixels.

Lenses

Plano-convex spherical lenses are available with focal lengths of 25, 50, 75, 100, 125, 150, 175, 200, and 250 mm. Plano-concave lenses with focal lengths of -30, -50, and -75 are also available. It is acceptable to mount a lens between the end of a tube and a tube ring or between two tube rings. To facilitate easy installation and removal, mount lenses in short (e.g. 0.5") lens tubes. In most cases, the convex side of the lens faces toward the collimated beam. The flat side goes toward the convergent rays.

Microscope design

There are a few additional things to keep in mind as you design your microscope. In traditional microscopes, the objective is above the sample and the observer looks down through an eyepiece. This is not a good arrangement for many biological applications. Looking through the growth medium in a petri dish or well distorts the image, so you would prefer to observe from the same side the cells are growing on. If you turn the dish upside down, all of to goo runs out on to the microscope stage. To solve this, biological microscopes usually place the objective underneath the sample. (This also explains why the print on the objective is upside down.) The optical path is pretty long, almost half a meter. It would be inconvenient to lie on the floor while making observations or stand on a ladder while changing samples, so most inverted microscopes use a 45 degree mirror, like M1 in the diagram below, to redirect light sideways.

20.309 microscope block diagram
Dichroic mirror

You are going to add fluorescence capability to your microscope in part 2. With a little bit of planning, you can build your part 1 ‘scope in such a way that you won't have to take your instrument apart for part 2. One of the key components needed for epifluorescence is a dichroic mirror, DM in the diagram. Dichroic mirrors reflect some colors of light, but pass others. The one you will use reflects the green laser light used to excite the sample but passes red light emitted by fluorescent molecules in the sample.

The transmission spectra for the 565DCXT dichroic and the E590LPv2 barrier filter

The dichroic mirror is imperfect and still allows a substantial amount of green light to pass. It is thus supplemented by a barrier filter BF to attenuate green light by 5 orders of magnitude. The combination DM + BF is essential for making good images.

In contrast with the transillumination bright-field approach where the LED incident light traverses the sample before reaching all image-forming optics, illumination for epi-fluorescence microscopy reaches the sample through the objective lens — from the same side of the sample that is observed. The rationale behind this design choice is the relative dimness of the fluorescence emission signal with respect to the needed excitation intensity. Even though fluorescence excitation and emission occur at distinct wavelengths and can thus be filtered away from one another, it is still advantageous to restrict to the utmost the amount of excitation light in the image plane.


In epi-fluorescence mode, the sample is illuminated by a green laser. Traveling through and focused by the objective lens, this collimated laser beam would result in single point illumination of the sample. To restore collimated excitation of a broader region of the sample, lens L5 must be inserted in the optical path... with the drawback that the laser beam's diameter, ~ 1.1 mm originally, thus becomes "minified" by a factor f5 / fobj, which would result in the illumination of a disc only ~ 25 μm in diameter! A Gallilean beam expander (L3 and L4) is thus the final requirement of the microscope design. It allows the collimated laser illumination to match the CCD camera field of view. The focal lengths chosen for L3 and L4, and thus the expansion factor of the beam expander, are a tradeoff between uniformity and brightness of illumination, given the Gaussian shape of the laser beam.

Optical microscopy lab

Code examples and simulations

Background reading

References

  1. Hooke, R. Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon London:Jo. Martyn, and Ja. Allestry, Printers to the Royal Society; 1665
  2. See, for example: McClintock, B. The origin and behavior of mutable loci in maize. PNAS. 1950; 36:344-355. [1], [2], and Endersby, Jim. A Guinea Pig's History of Biology. Cambridge, Massachusetts: Harvard University Press; 2007.
  3. See class stellar site for protocol. Oh & Yamaguchi, unpublished lab report