You will also be using fluorescence spectroscopy to assess the fluorescent properties of the cadmium sulfide particles you have collected with your yeast display model system.
 
You will also be using fluorescence spectroscopy to assess the fluorescent properties of the cadmium sulfide particles you have collected with your yeast display model system.
 +
 +
# Vortex to resuspend yeast pellet from M2D5.
 +
#*The yeast in this pellet were incubated with 100uM cadmium nitrate for 24 hours before being harvested through centrifugation.
 +
#*The media from this experiment was filtered and analyzed via ICP-OES.
 +
#Remove 5ul of yeast suspension and dilute it in 2ml dionized water.
 +
#Sonicate sample for 3 minutes to break apart yeast and dislodge cadmium sulfide particles.
 +
#*In your demonstration with Dr. Jifa Qi, you will only sonicate for 1 minute. This should be enough to see cadmium sulfide fluorescence
 +
#Filter your sample to remove yeast debris and collect cadmium sulfide in the supernatent.
 +
#Pipette 1ml of supernatent into a cuvette and read on the fluorimeter
 +
#*In the full-scale analysis of your cadmium sulfide particle production, Dr. Qi used a reusable quartz cuvette to measure lower wavelength excitation (250nm). By using the quartz cuvette, he was able to avoid interference by the polymers present in plastic disposable cuvettes.
 +
#*For your demonstration, you will use disposable cuvettes and a longer wavelength excitation (300-450nm). This will allow you to see the fluorescent shift, but isn't as accurate due to interference.
 +
#Working with Dr. Qi, adjust control parameters on the fluorimeter to allow for accurate capture and quantification of fluorescent emission.
 +
#*excitation wavelength
 +
#*fluorescence wavelength range
 +
#*measurement duration
  
 
===Part 3: Begin data analysis===
 
===Part 3: Begin data analysis===

Revision as of 13:10, 17 April 2024

20.109(S24): Laboratory Fundamentals of Biological Engineering

Sp24 banner image.png

Spring 2024 schedule        FYI        Assignments        Homework        Class data        Communication        Accessibility

       M1: Drug discovery        M2: Protein engineering        M3: Project design       


Introduction

The Transmission Electron Microscope (TEM) achieves its remarkable resolution by 'illuminating' samples using an electron beam in a vacuum rather than using a conventional light source in air. TEM enables 0.14 nm resolution. At this resolution, one can distinguish individual planes of atoms – making TEM an excellent materials characterization instrument. The visualization of single atoms allows one to determine many parameters that describe the structure of the material, such as crystal plane spacing. Additionally, TEMs can be equipped to perform energy-dispersive X-ray spectroscopy analysis (EDX/EDS/XEDS) or electron energy-loss spectroscopy analysis (EELS). The former, EDX, enables identification of the atomic composition of heavier elements in the material. EDX has limited detection of elements below sodium (Na) on the periodic table and no detection of elements below carbon (C). The latter, EELS, is most sensitive to lighter elements, typically from carbon to the 3d transition metals. EELS can reveal atomic composition, electronic properties of the valence and conduction band of a material, surface properties, and chemical bonding.

Although we will be using TEM for materials characterization, it should also be noted that TEM is used for the analysis of biological materials as well. Since the electron beam passes through the sample that is being examined, the sample must be sufficiently thin and sufficiently sturdy to be hit by electrons in a vacuum. Rather than imaging single atoms, biological samples display contrast of components via different densities. Areas of higher density show up as a darker color than areas of low density, creating a "shadow image." It is important to remember that many biological materials are damaged or destroyed by the incoming electrons and that the TEM can image only the species that survive this harsh treatment.

In TEM, the denser parts of the sample will absorb or scatter some of the electron beam, and it's the scattered electrons or those that pass through the sample that are focused using an electromagnetic lens. This "electron shadow" then strikes a fluorescent screen, giving rise to an image that varies in darkness according to the sample's density. For samples that are amenable to TEM, this form of examination can allow observation of ångström-sized objects and of cellular details down to near atomic levels.

The principles behind transmission electron microscopy are analogous to the principles behind an optical microscope. Instead of light and glass lenses, however, the TEM utilizes an electron beam focused via electromagnetic lenses. Since the TEM utilizes an electron beam rather than visible light, the microscope must also function under a vacuum. This condition allows the electrons to remain traveling in a straight line because they have no air particles with which to collide.

Specifically, at the top of the column of a TEM, a high voltage electron beam is emitted by a cathode and focused via a condenser lens. Then the beam is transmitted through the specimen. The resulting image is then magnified through a series of electromagnetic lenses until it is recorded by a fluorescent screen or camera.

Protocols

Part 1: Examining TEM images of cadmium sulfide particles

TEM grid with carbon mesh, photographed at 60X magnification
We will be imaging our cadmium sulfide particles via TEM. To prepare for TEM, your previously collected samples were dotted onto a very thin TEM grid.
  • TEM grids can be made of many kinds of materials. Each design, however, includes lines of a conductive metal which disperses the electron beam and thereby prevents destruction of the sample from the harsh beam.
  • Our TEM grids are composed of a nickel mesh with carbon mesh strung between the grids.

To prepare grids to image cadmium sulfide particles, the following steps were performed in the Belcher lab.

  1. Sonicate yeast to dislodge cadmium sulfide particles from cell surface
  2. Filter supernatent to remove remaining yeast debris.
  3. Vortex the filtered supernatent for at least 1 minute and immediately remove 10 μL of the cadmium sulfide suspension to place on the TEM grid that you have balanced in the tension tweezers.
    TEM grid balanced in tweezers

  4. Allow the cadmium sulfide particles to settle onto the grid undisturbed for 10 minutes.
  5. Remove the droplets of water from the grids by touching the edges with a Kimwipe, thereby wicking the solution off the grid.
  6. Wash the grids by adding 10 μL of sterile water onto each grid.
  7. Allow the grids to sit with water for 1-3 minutes and then wick dry.
  8. Wash the grids by adding 10 μL of 100% EtOH onto each grid.
  9. Allow the grids to sit with EtOH for 1-3 minutes and then wick dry, if it hasn't air dried.
  1. Once a sample has been applied to our TEM grids, we will only visualize the portions of the sample contacting the carbon mesh, along with any imperfections in the carbon mesh itself.

Today you will image the TEM grids that contain these cadmium sulfide particles. Due to space limitations, only 2 groups can join Prof. Belcher and Dr. Jifa Qi for the imaging at a time. Please bring computers to work on the Research Article during your waiting time.

Part 2: Examining cadmium sulfide particles with fluorescence spectroscopy

You will also be using fluorescence spectroscopy to assess the fluorescent properties of the cadmium sulfide particles you have collected with your yeast display model system.

  1. Vortex to resuspend yeast pellet from M2D5.
    • The yeast in this pellet were incubated with 100uM cadmium nitrate for 24 hours before being harvested through centrifugation.
    • The media from this experiment was filtered and analyzed via ICP-OES.
  2. Remove 5ul of yeast suspension and dilute it in 2ml dionized water.
  3. Sonicate sample for 3 minutes to break apart yeast and dislodge cadmium sulfide particles.
    • In your demonstration with Dr. Jifa Qi, you will only sonicate for 1 minute. This should be enough to see cadmium sulfide fluorescence
  4. Filter your sample to remove yeast debris and collect cadmium sulfide in the supernatent.
  5. Pipette 1ml of supernatent into a cuvette and read on the fluorimeter
    • In the full-scale analysis of your cadmium sulfide particle production, Dr. Qi used a reusable quartz cuvette to measure lower wavelength excitation (250nm). By using the quartz cuvette, he was able to avoid interference by the polymers present in plastic disposable cuvettes.
    • For your demonstration, you will use disposable cuvettes and a longer wavelength excitation (300-450nm). This will allow you to see the fluorescent shift, but isn't as accurate due to interference.
  6. Working with Dr. Qi, adjust control parameters on the fluorimeter to allow for accurate capture and quantification of fluorescent emission.
    • excitation wavelength
    • fluorescence wavelength range
    • measurement duration

Part 3: Begin data analysis

Flow Cytometry ICP-OES Fluorimetry TEM

Reagents

  • Cuvettes (VWR)
  • FluoroHub Fluorimeter (Horriba Scientific)
  • 400-mesh nickel Formvar grids (EMS) for TEM
  • JEOL2010F Electron Microscope