Line 17: Line 17:
 
==== 3/20/11 ====
 
==== 3/20/11 ====
 
Based on model system (blue LED + ND filter), I don't think gain is high enough. Increased gain of transimpedance amp from 2.5e4 to 2.5e5.
 
Based on model system (blue LED + ND filter), I don't think gain is high enough. Increased gain of transimpedance amp from 2.5e4 to 2.5e5.
 +
Well, it didn't help that much. So it's back the way it was. In any case, we get a significant signal (~0.25 units amplitude) with the model system. Huzzah!
  
 
==== 3/18/11 ====
 
==== 3/18/11 ====

Revision as of 23:51, 20 March 2011

Looking for weekly reports?

To-Do

  • research what to order for heated lid

Weekly goals

  • 3/25: Optics finished. Code started.
  • 4/1: Heated lid finished.
  • 4/8: Updated heat transfer function. Successful PCR.
  • 4/15: QRT-PCR code finished.
  • 4/22: Setup is ready for QRT-PCR. (i.e., the mechanical setup works)
  • 4/29: Successful QRT-PCR from the side.
  • 5/6: QRT-PCR from the top. Now which one was better?

Progress Notes

3/20/11

Based on model system (blue LED + ND filter), I don't think gain is high enough. Increased gain of transimpedance amp from 2.5e4 to 2.5e5. Well, it didn't help that much. So it's back the way it was. In any case, we get a significant signal (~0.25 units amplitude) with the model system. Huzzah!

3/18/11

Optical setup appears to be working.

3/16/11

Added dichroic mirror.

Question: how can I verify the optical setup? it seems to respond to driving signal, not fluorescence.

TODO: put away 25mm lens