In this exercise, you will figure out where stuff is in an image. | In this exercise, you will figure out where stuff is in an image. | ||
+ | |||
+ | ===Background=== | ||
+ | |||
+ | Before you begin this lab exercise, be certain you understand these concepts: | ||
+ | |||
+ | * [[Fluorescent probes]] | ||
+ | * [[Epifluorescence microscopy]] | ||
+ | * [[Optical detectors and noise]] | ||
+ | * [[Image acquisition and preprocessing]] | ||
+ | * [[Morphological processing and segmentation of fluorescent microsphere images]] | ||
===Goals=== | ===Goals=== | ||
− | * Develop software to | + | |
− | + | * Develop software to precisely locate fluorescent microspheres in a digital image | |
+ | * Understand the limit | ||
+ | |||
+ | ==Notation== | ||
+ | <center> | ||
+ | {| class="wikitalble" | ||
+ | |- | ||
+ | |<i>X<sub>i</sub></i> , <i>Y<sub>i</sub></i> | ||
+ | |Horizontal and vertical coordinates of the <i>i</i>th particle's centroid | ||
+ | |- | ||
+ | |<i>X<sub>i</sub> [n]</i> , <i>Y<sub>i</sub> [n]</i> | ||
+ | |Centroid of particle <i>i</i> in frame number <i>n</i> of an image sequence of <i>N</i> frames | ||
+ | |- | ||
+ | |<i>σ<sub>x,i</sub></i> , <i>σ<sub>y,i</sub> | ||
+ | |Standard deviation of particle <i>i</i> centroid in multiple images | ||
+ | |} | ||
+ | </center> | ||
==Procedure== | ==Procedure== |
Revision as of 21:34, 4 January 2013
{Template:20.309}
Overview
In this exercise, you will figure out where stuff is in an image.
Background
Before you begin this lab exercise, be certain you understand these concepts:
- Fluorescent probes
- Epifluorescence microscopy
- Optical detectors and noise
- Image acquisition and preprocessing
- Morphological processing and segmentation of fluorescent microsphere images
Goals
- Develop software to precisely locate fluorescent microspheres in a digital image
- Understand the limit
Notation
Xi , Yi | Horizontal and vertical coordinates of the ith particle's centroid |
Xi [n] , Yi [n] | Centroid of particle i in frame number n of an image sequence of N frames |
σx,i , σy,i | Standard deviation of particle <i>i centroid in multiple images |
Procedure
Image 1 micron fluorescent microspheres
- Use imaqtool to record an image of 1 micron fluorescent microspheres at 40X magnification.
- Minimize quantization noise by selecting 16-bit monochrome mode.
- Adjust the camera settings to achieve good contrast with minimal noise.
- Plot a histogram of the pixel values to verify optimal exposure
- Use a log axis for the pixel count.
- Take another image if the contrast is low or some pixels are over-saturated.
- Write a function in Matlab or another languageto compute the centroid of each microsphere
- The function should return an N row by 2 column matrix, where N is the number microspheres in the image. The first column should contain the X coordinate and the second column should contain Y.
Relevant Matlab commands
imaqtool imhist semilogy im2bw graythresh regionprops
Centroid precision and accuracy
- Image fluorescent microspheres as before, but change the settings in imaqtool to capture 5 sequential frames instead of 1.
- Ensure that the camera gain is set to 0.
- Run your centroid finding code on each image.
- Compute the standard deviation of each particle's centroid, σx and σy.
- Turn the gain up to 5; re-adjust the exposure setting, and acquire another set of 5 images
- Turn the gain up to 10; re-adjust the exposure setting, and acquire another set of 5 images
- Run your centroid finder on the gain=5 and gain=10 image sets and compute the standard deviation.
Report
- Compare the accuracy an precision of intensity-weighted centroids at different gain settings.
- How does the uncertainty depend on the exposure setting?
- What is the likely explanation of the trend?
Error sources
Discuss the effect of the following error sources on the centroid values.
- Shot noise
- Readout noise
- Dark current noise
- Ambient light
- Mechanical vibration
<-- Todo: image representations, histogram and how to expose image page imaqtool/camera settings page Determine gain value to use -->